MIRD Pamphlet No. 34, Part 2: Benchmarking of MIRDct Software for CT Organ Dose Estimation

Gunjan Kayal, Juan Camilo Ocampo Ramos, Laura E Dinwiddie, Jared M Baggett, Robert J Dawson, Harry Marquis, Stefan K Wehmeier, Oleksandra Ivashchenko, Lukas M Carter, Wesley E Bolch, Adam L Kesner

Journal of Nuclear Medicine · 2026

DOI: 10.2967/jnumed.125.271501 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Nuclear Medicine, 2026. Verfasst von Gunjan Kayal und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York; kayalg1@mskcc.org. Hauptthemen laut MEDLINE: Software, Tomography, X-Ray Computed, Radiation Dosage, Benchmarking. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate estimation of organ and effective doses in CT imaging is essential for risk assessment, protocol optimization, and personalized care in diagnostic radiology and nuclear medicine. We systematically benchmarked MIRDct, a freely available mesh phantom-based CT dose calculation software, against established reference software (National Cancer Institute Dosimetry System for Computed Tomography [NCICT] and VirtualDose [Virtual Phantoms]) by evaluating agreement between organ-absorbed doses and effective doses across representative scanners, phantoms, and protocols.

Methods

Organ absorbed and effective doses were calculated for adult and pediatric phantoms for whole-body (WB) and regional (head, chest, abdomen-pelvis [AP]) CT examinations. MIRDct uses mesh-based International Commission on Radiologic Protection (ICRP) reference phantoms with anatomically realistic organ surfaces, whereas NCICT and VirtualDose use voxel-based ICRP 110 and hybrid Rensselaer Polytechnic Institute/University of Florida phantom models, respectively. For each software, volumetric CT dose index (CTDIvol) values were obtained from the software interface using matched acquisition parameters; in MIRDct, these values were derived from scanner console-reported outputs for the corresponding protocol settings. Organ absorbed doses, dose coefficients, and effective doses were computed across 44 matched scanner-phantom-protocol configurations. Inter-software differences were summarized using medians and interquartile ranges. For regional protocols, organ-absorbed doses were stratified by irradiation category (in-field, partial-in-field, out-of-field), to assess field-dependent variability.

Results

CTDIvol​ values reported by the 3 software tools showed close agreement across matched protocol configurations, with median inter-software differences not exceeding 7%. For in-field organs, dose coefficients from NCICT and VirtualDose generally agreed with MIRDct values within ±25% across adult and pediatric head, chest, AP, and WB protocols, indicating good agreement in the primary beam region. Larger relative deviations occurred for partial-in-field and out-of-field organs, where doses were scatter-dominated; however, absolute organ doses were less than 2 mGy, limiting clinical relevance. Effective dose estimates showed similar concordance: differences were below 25% for all VirtualDose comparisons except head scans and for WB protocols, whereas adult chest and AP protocols differed by up to 40% relative to NCICT. These differences were associated with variations in phantom anatomy and fixed, pre-tabulated CTDIvol reference values in NCICT and VirtualDose, compared with protocol-specific, console-reported CTDIvol inputs in MIRDct.

Conclusion

MIRDct provides organ- and effective-dose estimates that are broadly consistent with established CT dosimetry tools, with agreement typically within ±25% for in-field organs and within a few milligray for absolute doses across adult and pediatric protocols. The use of mesh-based ICRP reference phantoms with anatomically realistic organ surfaces, protocol-specific CTDIvol inputs from the scanner console, and uncertainty propagation supports its application as a research tool for CT dose benchmarking, protocol optimization, and quality assurance in diagnostic CT and nuclear medicine.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SoftwareTomography, X-Ray ComputedRadiation DosageBenchmarkingRadiometryHumansPhantoms, ImagingAdultChild

Stichwörter der Autor:innen

CT dosesMIRDctcomputational phantomsdosimetry comparisoneffective doseorgan-absorbed dose

Autor:innen und Institutionen

  • Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York; kayalg1@mskcc.org.
  • Department of Imaging Physics, University of Texas M.D. Anderson Cancer Center, Houston, Texas.
  • J. Crayton Pruitt Department of Biomedical Engineering, University of Florida, Gainesville, Florida; and.
  • Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York.
  • Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, Groningen, The Netherlands.

Förderung

  • NCI NIH HHS
  • NIBIB NIH HHS

Zitieren

APA

Kayal, G., Ocampo Ramos, J. C., Dinwiddie, L. E., Baggett, J. M., Dawson, R. J., Marquis, H., Wehmeier, S. K., Ivashchenko, O., Carter, L. M., Bolch, W. E., & Kesner, A. L. (2026). MIRD Pamphlet No. 34, Part 2: Benchmarking of MIRDct Software for CT Organ Dose Estimation. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.125.271501

Vancouver

Kayal G, Ocampo Ramos JC, Dinwiddie LE, Baggett JM, Dawson RJ, Marquis H, et al. MIRD Pamphlet No. 34, Part 2: Benchmarking of MIRDct Software for CT Organ Dose Estimation. Journal of Nuclear Medicine. 2026. doi:10.2967/jnumed.125.271501

BibTeX
@article{Kayal2026,
  author = {Kayal, Gunjan and Ocampo Ramos, Juan Camilo and Dinwiddie, Laura E and Baggett, Jared M and Dawson, Robert J and Marquis, Harry and Wehmeier, Stefan K and Ivashchenko, Oleksandra and Carter, Lukas M and Bolch, Wesley E and Kesner, Adam L},
  title = {MIRD Pamphlet No. 34, Part 2: Benchmarking of MIRDct Software for CT Organ Dose Estimation},
  journal = {Journal of Nuclear Medicine},
  year = {2026},
  doi = {10.2967/jnumed.125.271501}
}
Kostenlos starten →