Task-based automatic keV selection: leveraging routine virtual monoenergetic imaging for dose reduction on clinical photon-counting detector CT *
Physics in Medicine & Biology · 2024
DOI: 10.1088/1361-6560/ad41b3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2024. Verfasst von Kishore Rajendran und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, MN, United States of America. Hauptthemen laut MEDLINE: Radiation Dosage, Photons, Phantoms, Imaging, Tomography, X-Ray Computed. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Objective . Photon-counting detector (PCD) CT enables routine virtual-monoenergetic image (VMI) reconstruction. We evaluated the performance of an automatic VMI energy level (keV) selection tool on a clinical PCD-CT system in comparison to an automatic tube potential (kV) selection tool from an energy-integrating-detector (EID) CT system from the same manufacturer. Approach. Four torso-shaped phantoms (20–50 cm width) containing iodine (2, 5, and 10 mg cc −1 ) and calcium (100 mg cc −1 ) were scanned on PCD-CT and EID-CT. Dose optimization techniques, task-based VMI energy level and tube-potential selection on PCD-CT (CARE keV) and task-based tube potential selection on EID-CT (CARE kV), were enabled. CT numbers, image noise, and dose-normalized contrast-to-noise ratio (CNR d ) were compared. Main results . PCD-CT produced task-specific VMIs at 70, 65, 60, and 55 keV for non-contrast, bone, soft tissue with contrast, and vascular settings, respectively. A 120 kV tube potential was automatically selected on PCD-CT for all scans. In comparison, EID-CT used x-ray tube potentials from 80 to 150 kV based on imaging task and phantom size. PCD-CT achieved consistent dose reduction at 9%, 21% and 39% for bone, soft tissue with contrast, and vascular tasks relative to the non-contrast task, independent of phantom size. On EID-CT, dose reduction factor for contrast tasks relative to the non-contrast task ranged from a 65% decrease (vascular task, 70 kV, 20 cm phantom) to a 21% increase (soft tissue with contrast task, 150 kV, 50 cm phantom) due to size-specific tube potential adaptation. PCD-CT CNR d was equivalent to or higher than those of EID-CT for all tasks and phantom sizes, except for the vascular task with 20 cm phantom, where 70 kV EID-CT CNR d outperformed 55 keV PCD-CT images. Significance . PCD-CT produced more consistent CT numbers compared to EID-CT due to standardized VMI output, which greatly benefits standardization efforts and facilitates radiation dose reduction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Mayo Clinic, Rochester, MN, United States of America.
ORCID-Profile: Kishore Rajendran
Förderung
- NIBIB NIH HHS
- National Institutes of Health
Zitieren
APA
Rajendran, K., Bruesewitz, M., Swicklik, J., Ferrero, A., Thorne, J., Yu, L., McCollough, C., & Leng, S. (2024). Task-based automatic keV selection: leveraging routine virtual monoenergetic imaging for dose reduction on clinical photon-counting detector CT *. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ad41b3
Vancouver
Rajendran K, Bruesewitz M, Swicklik J, Ferrero A, Thorne J, Yu L, et al. Task-based automatic keV selection: leveraging routine virtual monoenergetic imaging for dose reduction on clinical photon-counting detector CT *. Physics in Medicine & Biology. 2024. doi:10.1088/1361-6560/ad41b3
BibTeX
@article{Rajendran2024,
author = {Rajendran, Kishore and Bruesewitz, Michael and Swicklik, Joseph and Ferrero, Andrea and Thorne, Jamison and Yu, Lifeng and McCollough, Cynthia and Leng, Shuai},
title = {Task-based automatic keV selection: leveraging routine virtual monoenergetic imaging for dose reduction on clinical photon-counting detector CT *},
journal = {Physics in Medicine & Biology},
year = {2024},
doi = {10.1088/1361-6560/ad41b3}
}
EUVIMED