Radiation dose and signal-to-noise ratio in pediatric head CT: a phantom study comparing photon-counting and energy-integrating detectors

Laura Valentina Klüner, Sebastian Zensen, Hannah Peuster, Raya Ocker-Serger, Marcel Drews, Denise Bos, Cornelius Deuschl, Lale Umutlu, Michael Forsting, Johannes Haubold, Marcel Opitz

Neuroradiology · 2026

DOI: 10.1007/s00234-026-04075-9 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Neuroradiology, 2026. Verfasst von Laura Valentina Klüner und 10 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Diagnostic and Interventional Radiology and Neuroradiology, Essen University Hospital, Essen, Germany. laura.kluener@uk-essen.de. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed, Head. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Essen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study evaluated the relationship between image quality level (IQL), radiation dose, image noise, and signal-to-noise ratio (SNR) in pediatric head CT, comparing energy-integrating detector CT (EID-CT) and photon-counting CT (PCCT) across age-specific phantoms.

Methods

Head CT scans of 1-, 5-, and 10-year-old anthropomorphic phantoms were performed across a range of IQLs on EID-CT and PCCT. For each IQL, radiation dose, image noise, and SNR were recorded and compared between the two systems.

Results

PCCT required lower CTDI vol across all phantoms ( p < 0.05). In the clinically relevant IQL-range (IQL 200–300) the 10-year-old phantom showed the greatest relative radiation dose reduction using PCCT (9%), compared to 0.45% in the 1-year-old phantom. Organ doses were highest in the 1-year-old phantom for red bone marrow and skin with no significant differences between the two systems. PCCT offered significantly higher SNR than EID-CT for brain parenchyma ( p < 0.03) and bone ( p = 0.012). Image noise was significantly lower in PCCT for brain parenchyma in the 1- and 5-year-old phantoms ( p < 0.03) and for bone in the 1-year-old phantom ( p = 0.012). Over the full IQL-range (1-300) the average SNR was up to ~ 20% higher for brain parenchyma and ~ 50% higher for bone with PCCT than with EID-CT for all phantoms.

Conclusion

PCCT offers superior SNR for brain and bone while reducing radiation dose, with the largest radiation dose benefit in older pediatric phantoms, supporting optimized pediatric neuroimaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosageTomography, X-Ray ComputedHeadHumansPhantoms, ImagingSignal-To-Noise RatioChildInfantChild, PreschoolPhotons

Stichwörter der Autor:innen

EID-CTHeadImage qualityPediatricPhoton-counting CTSNR

Autor:innen und Institutionen

  • Institute of Diagnostic and Interventional Radiology and Neuroradiology, Essen University Hospital, Essen, Germany. laura.kluener@uk-essen.de.
  • Institute of Diagnostic and Interventional Radiology and Neuroradiology, Essen University Hospital, Essen, Germany.
  • Faculty of Medicine, University of Duisburg-Essen, Essen, Germany.
  • Institute of Diagnostic and Interventional Radiology, University Hospital of Zurich, Zurich, Switzerland.

Förderung

  • Universitätsklinikum Essen

Zitieren

APA

Klüner, L. V., Zensen, S., Peuster, H., Ocker-Serger, R., Drews, M., Bos, D., Deuschl, C., Umutlu, L., Forsting, M., Haubold, J., & Opitz, M. (2026). Radiation dose and signal-to-noise ratio in pediatric head CT: a phantom study comparing photon-counting and energy-integrating detectors. Neuroradiology. https://doi.org/10.1007/s00234-026-04075-9

Vancouver

Klüner LV, Zensen S, Peuster H, Ocker-Serger R, Drews M, Bos D, et al. Radiation dose and signal-to-noise ratio in pediatric head CT: a phantom study comparing photon-counting and energy-integrating detectors. Neuroradiology. 2026. doi:10.1007/s00234-026-04075-9

BibTeX
@article{Klner2026,
  author = {Klüner, Laura Valentina and Zensen, Sebastian and Peuster, Hannah and Ocker-Serger, Raya and Drews, Marcel and Bos, Denise and Deuschl, Cornelius and Umutlu, Lale and Forsting, Michael and Haubold, Johannes and Opitz, Marcel},
  title = {Radiation dose and signal-to-noise ratio in pediatric head CT: a phantom study comparing photon-counting and energy-integrating detectors},
  journal = {Neuroradiology},
  year = {2026},
  doi = {10.1007/s00234-026-04075-9}
}
Kostenlos starten →