Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A phantom study

Joël Greffier, Djamel Dabli, Sebastian Faby, Maxime Pastor, Cédric Croisille, Fabien de Oliveira, Julien Erath, Jean Paul Beregi

Diagnostic and Interventional Imaging · 2024

DOI: 10.1016/j.diii.2024.05.002 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Diagnostic and Interventional Imaging, 2024. Verfasst von Joël Greffier und 7 weiteren Autor:innen. Beteiligte Einrichtung: IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France. Electronic address: joel.gr…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiation Dosage, Tomography, X-Ray Computed, Radiography, Abdominal. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 37 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The purpose of this study was to assess image-quality and dose reduction potential using a photon-counting computed tomography (PCCT) system by comparison with two different dual-source CT (DSCT) systems using two phantoms. MATERIALS AND METHODS: Acquisitions on phantoms were performed using two DSCT systems (DSCT1 [Somatom Force] and DSCT2 [Somatom Pro.Pulse]) and one PCCT system (Naeotom Alpha) at four dose levels (13/6/3.4/1.8 mGy). Noise power spectrum (NPS) and task-based transfer function (TTF) were computed to assess noise magnitude and noise texture and spatial resolution (f50), respectively. Detectability indexes (d') were computed to model the detection of abdominal lesions: one unenhanced high-contrast task, one contrast-enhanced high-contrast task and one unenhanced low-contrast task. Image quality was subjectively assessed on an anthropomorphic phantom by two radiologists. RESULTS: For all dose levels, noise magnitude values were lower with PCCT than with DSCTs. For all CT systems, similar noise texture values were found at 13 and 6 mGy, but the greatest noise texture values were found for DSCT2 and the lowest for PCCT at 3.4 and 1.8 mGy. For high-contrast inserts, similar or lower f50 values were found with PCCT than with DSCT1 and the opposite pattern was found for the low-contrast insert. For the three simulated lesions, d' values were greater with PCCT than with DSCTs. Abdominal images were rated satisfactory for clinical use by the radiologists for all dose levels with PCCT and for 13 and 6 mGy with DSCTs. CONCLUSION: By comparison with DSCTs, PCCT reduces image-noise and improves detectability of simulated abdominal lesions without altering the spatial resolution and image texture. Image-quality obtained with PCCT seem to indicate greater potential for dose optimization than those obtained with DSCTs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiation DosageTomography, X-Ray ComputedRadiography, AbdominalPhotonsHumansSignal-To-Noise Ratio

Stichwörter der Autor:innen

AbdomenIterative reconstructionMultidetector computed tomographyPhoton counting CTTask-based image quality assessment

Autor:innen und Institutionen

  • IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France. Electronic address: joel.greffier@chu-nimes.fr.
  • IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
  • Department of Computed Tomography, Siemens Healthineers AG, 91301 Forchheim, Germany.

Zitieren

APA

Greffier, J., Dabli, D., Faby, S., Pastor, M., Croisille, C., de Oliveira, F., Erath, J., & Beregi, J. P. (2024). Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A phantom study. Diagnostic and Interventional Imaging. https://doi.org/10.1016/j.diii.2024.05.002

Vancouver

Greffier J, Dabli D, Faby S, Pastor M, Croisille C, de Oliveira F, et al. Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A phantom study. Diagnostic and Interventional Imaging. 2024. doi:10.1016/j.diii.2024.05.002

BibTeX
@article{Greffier2024,
  author = {Greffier, Joël and Dabli, Djamel and Faby, Sebastian and Pastor, Maxime and Croisille, Cédric and de Oliveira, Fabien and Erath, Julien and Beregi, Jean Paul},
  title = {Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A phantom study},
  journal = {Diagnostic and Interventional Imaging},
  year = {2024},
  doi = {10.1016/j.diii.2024.05.002}
}
Kostenlos starten →