Three‐dimensional noise characteristics of clinical photon counting detector CT
Medical Physics · 2025
DOI: 10.1002/mp.70067 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Humberto Monsivais und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Photons, Tomography, X-Ray Computed, Signal-To-Noise Ratio, Imaging, Three-Dimensional. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Since the introduction of whole‐body photon‐counting detector CT (PCD‐CT) into clinical practice, extensive physics assessments have been conducted to elucidate its image quality advantages over energy‐integrating detector CT (EID‐CT) and to support its clinical adoption. However, evaluations of its three‐dimensional (3D) noise power spectrum (NPS), which simultaneously quantifies in‐plane and through‐plane noise texture and magnitude, remain limited.
Purpose
To experimentally evaluate the 3D NPS of an NAEOTOM Alpha PCD‐CT system and its dependence on scan mode, reconstruction image type, quantum iterative reconstruction (QIR) strength, mono‐energy (keV) level, spiral pitch, and radiation dose.
Methods
Repeated scans of a 20 cm water phantom and a 30 cm PMMA phantom were conducted under the clinical Standard mode, clinical Ultra‐High‐Resolution (UHR) mode, and an Expert Service mode. Reconstructed image types include T3D, virtual monoenergetic image (VMI), and T1 (a linear reconstruction of total‐energy bin available via the Expert Service mode). Data were collected at seven dose levels (0.4–24 mGy) and four spiral pitch levels (0.35–1.5). T3D and VMI images were reconstructed with varying QIR strengths, and VMIs were reconstructed at energies ranging from 40 to 190 keV. The 3D NPS, , was calculated from each ensemble of 3D image volumes. Axial was obtained by integrating NPS 3D along , while was obtained by integrating NPS 3D over and . Results of T1 images were flat, indicating no noise correlation across PCD rows. In contrast, all clinical‐mode reconstructions exhibited through‐plane noise correlation, as reflected in the shape of their . For clinical‐mode reconstructions, the shape of their 3D NPS showed a mild to moderate dependence on dose, with lower doses producing NPS profiles shifted towards lower frequencies in both axial and z directions. With a matched post‐object radiation exposure, the larger phantom resulted in higher noise and stronger noise correlation compared to the smaller phantom. QIR only mildly enhanced noise correlation. For a given CTDI vol , spiral pitch has a negligible impact on 3D NPS. Due to through‐plane noise correlation, the variance of T3D and VMI decreases with slice thickness () approximately as , in contrast to the scaling observed in T1 images. The shape of the 3D NPS of VMI showed only weak dependence on the keV level along the axial frequency direction. Compared to the Standard mode, the UHR mode reduced image variance by 26% when using a soft‐tissue (Br44) kernel and by 77% with a sharp (Br76) kernel. However, 3D NPS analysis revealed stronger through‐plane noise correlation in UHR images.
Conclusion
The 3D NPS provides new insight into the noise characteristics of PCD‐CT: Noise in the native PCD‐CT projection data is uncorrelated across detector rows, but clinical reconstruction processes introduce noise spatial correlation along both axial and z directions, particularly at higher QIR strengths, lower radiation doses, or with larger image objects that increase the percentage of scattered photons. Compared to the Standard mode, UHR mode reconstruction exhibits stronger noise correlation due to its superior detector spatial resolution, allowing for more aggressive spatial smoothing to achieve the desired spatial resolution in the final image.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Abdominal Imaging, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Interventional Radiology, University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
Zitieren
APA
Monsivais, H., Liu, X., Dong, F., Jacobsen, M. C., Rong, J., Jensen, C. T., & Li, K. (2025). Three‐dimensional noise characteristics of clinical photon counting detector CT. Medical Physics. https://doi.org/10.1002/mp.70067
Vancouver
Monsivais H, Liu X, Dong F, Jacobsen MC, Rong J, Jensen CT, et al. Three‐dimensional noise characteristics of clinical photon counting detector CT. Medical Physics. 2025. doi:10.1002/mp.70067
BibTeX
@article{Monsivais2025,
author = {Monsivais, Humberto and Liu, Xinming and Dong, Frank and Jacobsen, Megan C and Rong, John and Jensen, Corey T and Li, Ke},
title = {Three‐dimensional noise characteristics of clinical photon counting detector CT},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70067}
}
EUVIMED