RF safety assessment of dental retainers for human head MRI array coils at 9.4 T

Egor V Berezko, Georgiy A Solomakha, Nikolai I Avdievich, Klaus Scheffler

Magnetic Resonance Imaging · 2027-01 · 135 · 110803

DOI: 10.1016/j.mri.2026.110803 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2027-01. Verfasst von Egor V Berezko und 3 weiteren Autor:innen. Beteiligte Einrichtung: High-field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany. Electronic address: egor.berezko@tuebingen.…. Die Arbeit nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council, German Research Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this work we developed a numerical workflow for evaluating radiofrequency (RF) safety for ultra-high field (UHF, 7 T and above) MRI with a local transmit array coil in the presence of conductive objects such as dental retainers. The proposed workflow includes evaluation of the B1+ distribution and peak Specific Absorption Rate (pSAR) for the standard circularly polarized modes, worst-case pSAR modes, and random RF modes. The workflow provides a safety margin, which limits the RF transmit power during MRI scanning. The proposed workflow is applicable to different array coils at different UHF strengths (e.g. 7 T, 9.4 T, 11.7 T, 14 T) as well as to other types of conductive objects (e.g., dental or neural implants). 16-element tight-fit transceiver and larger 16-element transmit-only head arrays were numerically evaluated at 9.4 T in the presence of the dental retainer. Updated safety margins were calculated for both arrays. In addition, we calculated Q-matrix and virtual observation point matrices, which facilitate the real-time SAR-monitoring and can be used for optimization of the B1+ homogeneity and SAR.

Schlagwörter

Stichwörter der Autor:innen

9.4 TeslaDental retainersNumerical simulationRF safetySpecific absorption rateVOP compression

Autor:innen und Institutionen

  • High-field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany. Electronic address: egor.berezko@tuebingen.mpg.de.
  • High-field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • High-field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany; Department for Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany.

Förderung

  • European Research Council
  • German Research Foundation

Zitieren

APA

Berezko, E. V., Solomakha, G. A., Avdievich, N. I., & Scheffler, K. (2027). RF safety assessment of dental retainers for human head MRI array coils at 9.4 T. Magnetic Resonance Imaging, 135, 110803. https://doi.org/10.1016/j.mri.2026.110803

Vancouver

Berezko EV, Solomakha GA, Avdievich NI, Scheffler K. RF safety assessment of dental retainers for human head MRI array coils at 9.4 T. Magnetic Resonance Imaging. 2027;135:110803. doi:10.1016/j.mri.2026.110803

BibTeX
@article{Berezko2027,
  author = {Berezko, Egor V and Solomakha, Georgiy A and Avdievich, Nikolai I and Scheffler, Klaus},
  title = {RF safety assessment of dental retainers for human head MRI array coils at 9.4 T},
  journal = {Magnetic Resonance Imaging},
  year = {2027},
  volume = {135},
  pages = {110803},
  doi = {10.1016/j.mri.2026.110803}
}
Kostenlos starten →