EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Quality assurance for intraoperative MRI RF coils in stereotactic neurosurgical planning

James C Thorpe iD, Rafal Panek iD, Charlotte Bull iD, Paul S Morgan iD

Journal of Applied Clinical Medical Physics · 2026-07 · 27 (7)

DOI: 10.1002/acm2.70699 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 15.07.2026. Verfasst von James C Thorpe und 3 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics, Nottingham University Hospitals NHS Trust, Nottingham, UK. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging, Quality Assurance, Health Care, Brain Neoplasms. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The application of MRI during a surgical procedure (intraoperative MRI) provides the surgical team with up‐to‐date anatomical information to improve precision and inform decision making during the operation. This requires the use of a stereotactic frame or RF coil that affixes to the patient throughout the procedure, with the clinical team requiring assurance that the image quality is consistent and appropriate. The following QA program has been developed in order to monitor the condition of an intraoperative head coil.

Methods

This was set up for a third party intraoperative 8‐channel head coil in conjunction with a 3.0T MRI scanner. QA testing consists of acquiring a large Field of View isometric (1.5 mm) 3D T1‐weighted rapid gradient echo (FFE) of a 5000 CC Phantom Bottle, with two dynamic acquisitions and the individual channel data saved. Offline analysis is then performed to calculate the signal‐to‐noise ratio (SNR) of the phantom within the coil, the SNR for each channel, and the distance (D) from the coil at which the SNR decreases to half the value within the coil.

Results

Testing has been performed on a monthly basis since May 2022 with consistent values for SNR (including for each channel) and D. There were three occasions where the coil failed the QA tests.

Conclusion

A monthly QA program has been developed to monitor coil element performance, confirming that the coil has been functioning as expected.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingPhantoms, ImagingQuality Assurance, Health CareBrain NeoplasmsRadiosurgeryRadiotherapy Planning, Computer-AssistedNeurosurgical ProceduresHumansSignal-To-Noise RatioImage Processing, Computer-Assisted

Stichwörter der Autor:innen

MRIdecision makingimagingneurosurgeryquality assurance

Autor:innen und Institutionen

  • Medical Physics, Nottingham University Hospitals NHS Trust, Nottingham, UK.
  • School of Medicine, University Of Nottingham, Nottingham, UK.

ORCID-Profile: James C Thorpe, Rafal Panek, Charlotte Bull, Paul S Morgan

Zitieren

APA

Thorpe, J. C., Panek, R., Bull, C., & Morgan, P. S. (2026). Quality assurance for intraoperative MRI RF coils in stereotactic neurosurgical planning. Journal of Applied Clinical Medical Physics, 27(7). https://doi.org/10.1002/acm2.70699

Vancouver

Thorpe JC, Panek R, Bull C, Morgan PS. Quality assurance for intraoperative MRI RF coils in stereotactic neurosurgical planning. Journal of Applied Clinical Medical Physics. 2026;27(7). doi:10.1002/acm2.70699

BibTeX
@article{Thorpe2026,
  author = {Thorpe, James C and Panek, Rafal and Bull, Charlotte and Morgan, Paul S},
  title = {Quality assurance for intraoperative MRI RF coils in stereotactic neurosurgical planning},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {7},
  doi = {10.1002/acm2.70699}
}
Kostenlos starten →