First-in-Human Intratumoral Temperature Monitoring During Standard 3 T MRI Demonstrates RF-Induced Tissue Heating Within Clinical Safety Limits
Bioengineering · 28.06.2026 · 13 (7) · 756
DOI: 10.3390/bioengineering13070756 ↗Kurz gefasst
Fallbericht in der Zeitschrift Bioengineering, online erschienen am 28.06.2026. Verfasst von Chie-Hee Cho und 12 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Diagnostic and Interventional Radiology, Jena University Hospital, Friedrich Schiller University Jena, 07747 Jena, Germany. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Magnetic resonance imaging (MRI) uses radiofrequency (RF) energy to generate diagnostic images. RF–tissue interactions lead to energy absorption and tissue heating, quantified by the specific absorption rate (SAR). Although SAR limits are strictly regulated for patient safety, actual in vivo tissue temperature changes during clinical MRI examinations in humans have not been directly measured. A patient with a histologically confirmed soft tissue sarcoma of the thigh underwent a clinically indicated 3 T MRI examination 24 h prior to resection. During imaging with whole-body SAR of 2.27 W/kg, direct temperature measurements (invasive and on the skin) were obtained. Temperatures increased by 2.0 °C within the tumor and at the skin surface was 3.4 °C at the skin surface. No technical difficulties or adverse events were observed, and the patient tolerated the examination well. This first-in-human case demonstrates the feasibility and safety of direct intratumoral temperature measurement during standard 3T MRI. While MRI was performed within safety limits of SAR as a surrogate for true tissue temperature, non-invasive temperature monitoring during MRI needs improvement. Controlled RF-induced heating during MRI may open new therapeutic possibilities, including MR-guided hyperthermia for sarcomas and other solid tumors or modulation of blood–brain barrier through transient RF-induced temperature elevations facilitating drug delivery.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Diagnostic and Interventional Radiology, Jena University Hospital, Friedrich Schiller University Jena, 07747 Jena, Germany.
- Department of Medical Physics, Clinic for Radiationtherapy, University Hospital Essen, 45147 Essen, Germany.
- Experimental Radiology, Jena University Hospital, 07747 Jena, Germany.
- Experimental Traumatology, Jena University Hospital, 07747 Jena, Germany.
- Department of Trauma, Hand and Reconstructive Surgery and Orthopedics, Clinic for Trauma, Jena University Hospital, 07747 Jena, Germany.
- Section of Pathology, Institute for Forensic Medicine, Jena University Hospital, 07747 Jena, Germany.
ORCID-Profile: Wolfram Weschenfelder, Ingrid Hilger, Britt Wildemann
Zitieren
APA
Cho, C. H., Bergholz, F., Lüdemann, L., Bergholz, C., Winger, E., Brand, P., Spiegel, C., Weschenfelder, W., Gaßler, N., Xylander, A., Hilger, I., Wildemann, B., & Hofmann, G. O. (2026). First-in-Human Intratumoral Temperature Monitoring During Standard 3 T MRI Demonstrates RF-Induced Tissue Heating Within Clinical Safety Limits. Bioengineering, 13(7), 756. https://doi.org/10.3390/bioengineering13070756
Vancouver
Cho CH, Bergholz F, Lüdemann L, Bergholz C, Winger E, Brand P, et al. First-in-Human Intratumoral Temperature Monitoring During Standard 3 T MRI Demonstrates RF-Induced Tissue Heating Within Clinical Safety Limits. Bioengineering. 2026;13(7):756. doi:10.3390/bioengineering13070756
BibTeX
@article{Cho2026,
author = {Cho, Chie-Hee and Bergholz, Franz and Lüdemann, Lutz and Bergholz, Carlo and Winger, Emma and Brand, Pauline and Spiegel, Christian and Weschenfelder, Wolfram and Gaßler, Nikolaus and Xylander, Anna and Hilger, Ingrid and Wildemann, Britt and Hofmann, Gunther O},
title = {First-in-Human Intratumoral Temperature Monitoring During Standard 3 T MRI Demonstrates RF-Induced Tissue Heating Within Clinical Safety Limits},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {7},
pages = {756},
doi = {10.3390/bioengineering13070756}
}