Commissioning Intracranial Stereotactic Radiosurgery for a Magnetic Resonance-Guided Radiation Therapy (MRgRT) System: MR-RT Localization and Dosimetric End-to-End Validation
International Journal of Radiation Oncology*Biology*Physics · 2024
DOI: 10.1016/j.ijrobp.2023.08.043 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2024. Verfasst von Kathryn E Mittauer und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida; Herbert Wertheim College of Medicine, Florida…. Hauptthemen laut MEDLINE: Radiosurgery. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: ViewRay.
Abstract (Originalsprache)
PURPOSE: This is the first reporting of the MRIdian A3iTM intracranial package (BrainTxTM) and benchmarks the end-to-end localization and dosimetric accuracy for commissioning an magnetic resonace (MR)-guided stereotactic radiosurgery program. We characterized the localization accuracy between MR and radiation (RT) isocenter through an end-to-end hidden target test, relative dose profile intercomparison, and absolute dose validation. METHODS AND MATERIALS: BrainTx consists of a dedicated head coil, integrated mask immobilization system, and high-resolution MR sequences. Coil and baseplate attenuation was quantified. An in-house phantom (Cranial phantOm foR magNetic rEsonance Localization of a stereotactIc radiosUrgery doSimeter, CORNELIUS) was developed from a mannequin head filled with silicone gel, film, and MR BB with pinprick. A hidden target test evaluated MR-RT localization of the 1×1×1 mm3 TrueFISP MR and relative dose accuracy in film for a 1 cm diameter (International Electrotechnical Commission (IEC)-X/IEC-Y) and 1.5 cm diameter (IEC-Y/IEC-Z) spherical target. Two clinical cases (irregular-shaped target and target abutting brainstem) were mapped to the CORNELIUS phantom for feasibility assessment. A 2-dimensional (2D)-gamma compared calculated and measured dose for spherical and clinical targets with 1 mm/1% and 2 mm/2% criteria, respectively. A small-field chamber (A26MR) measured end-to-end absolute dose for a 1 cm diameter target. RESULTS: Coil and baseplate attenuation were 0.7% and 2.7%, respectively. The displacement of MR to RT localization as defined through the pinprick was 0.49 mm (IEC-X), 0.27 mm (IEC-Y), and 0.51 mm (IEC-Z) (root mean square 0.76 mm). The reproducibility across IEC-Y demonstrated high fidelity (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida; Herbert Wertheim College of Medicine, Florida International University, Miami, Florida. Electronic address: kathryn…
- Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, Florida; Herbert Wertheim College of Medicine, Florida International University, Miami, Florida.
Förderung
- ViewRay
Zitieren
APA
Mittauer, K. E., Tolakanahalli, R., Kotecha, R., Chuong, M. D., Mehta, M. P., Gutierrez, A. N., & Bassiri, N. (2024). Commissioning Intracranial Stereotactic Radiosurgery for a Magnetic Resonance-Guided Radiation Therapy (MRgRT) System: MR-RT Localization and Dosimetric End-to-End Validation. International Journal of Radiation Oncology*Biology*Physics. https://doi.org/10.1016/j.ijrobp.2023.08.043
Vancouver
Mittauer KE, Tolakanahalli R, Kotecha R, Chuong MD, Mehta MP, Gutierrez AN, et al. Commissioning Intracranial Stereotactic Radiosurgery for a Magnetic Resonance-Guided Radiation Therapy (MRgRT) System: MR-RT Localization and Dosimetric End-to-End Validation. International Journal of Radiation Oncology*Biology*Physics. 2024. doi:10.1016/j.ijrobp.2023.08.043
BibTeX
@article{Mittauer2024,
author = {Mittauer, Kathryn E and Tolakanahalli, Ranjini and Kotecha, Rupesh and Chuong, Michael D and Mehta, Minesh P and Gutierrez, Alonso N and Bassiri, Nema},
title = {Commissioning Intracranial Stereotactic Radiosurgery for a Magnetic Resonance-Guided Radiation Therapy (MRgRT) System: MR-RT Localization and Dosimetric End-to-End Validation},
journal = {International Journal of Radiation Oncology*Biology*Physics},
year = {2024},
doi = {10.1016/j.ijrobp.2023.08.043}
}
EUVIMED