MRI compatibility testing of commercial high intensity focused ultrasound transducers
Physica Medica · 2024
DOI: 10.1016/j.ejmp.2023.103194 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physica Medica, 2024. Verfasst von Nikolas Evripidou und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Transducers. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 43 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: The study aimed to compare the performance of eight commercially available single-element High Intensity Focused Ultrasound (HIFU) transducers in terms of Magnetic Resonance Imaging (MRI) compatibility. METHODS: Imaging of an agar-based MRI phantom was performed in a 3 T MRI scanner utilizing T2-Weighted Fast Spin Echo (FSE) and Fast low angle shot (FLASH) sequences, which are typically employed for high resolution anatomical imaging and thermometry, respectively. Reference magnitude and phase images of the phantom were compared with images acquired in the presence of each transducer in terms of the signal to noise ratio (SNR), introduced artifacts, and overall image quality. RESULTS: The degree of observed artifacts highly differed among the various transducers. The transducer whose backing material included magnetic impurities showed poor performance in the MRI, introducing significant susceptibility artifacts such as geometric distortions and signal void bands. Additionally, it caused the most significant SNR drop. Other transducers were shown to exhibit high level of MRI compatibility as the resulting images closely resembled the reference images with minimal to no apparent artifacts and comparable SNR values. CONCLUSIONS: The study findings may facilitate researchers to select the most suitable transducer for their research, simultaneously avoiding unnecessary testing. The study further provides useful design considerations for MRI compatible transducers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
- Department of Interventional Radiology, German Oncology Center, Limassol, Cyprus.
- VET EX MACHINA Limited, Nicosia, Cyprus.
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Electronic address: christakis.damianou@cut.ac.cy.
Zitieren
APA
Evripidou, N., Antoniou, A., Georgiou, L., Ioannides, C., Spanoudes, K., & Damianou, C. (2024). MRI compatibility testing of commercial high intensity focused ultrasound transducers. Physica Medica. https://doi.org/10.1016/j.ejmp.2023.103194
Vancouver
Evripidou N, Antoniou A, Georgiou L, Ioannides C, Spanoudes K, Damianou C. MRI compatibility testing of commercial high intensity focused ultrasound transducers. Physica Medica. 2024. doi:10.1016/j.ejmp.2023.103194
BibTeX
@article{Evripidou2024,
author = {Evripidou, Nikolas and Antoniou, Anastasia and Georgiou, Leonidas and Ioannides, Cleanthis and Spanoudes, Kyriakos and Damianou, Christakis},
title = {MRI compatibility testing of commercial high intensity focused ultrasound transducers},
journal = {Physica Medica},
year = {2024},
doi = {10.1016/j.ejmp.2023.103194}
}
EUVIMED