Increasing the scan‐efficiency of pulmonary imaging at 0. 55 T using iterative concomitant field and motion‐corrected reconstruction
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30054 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Ahsan Javed und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland,…. Hauptthemen laut MEDLINE: Lung, Magnetic Resonance Imaging. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Intramural NIH HHS, Division of Intramural Research.
Abstract (Originalsprache)
Purpose
To develop an iterative concomitant field and motion corrected (iCoMoCo) reconstruction for isotropic high‐resolution UTE pulmonary imaging at 0.55 T.
Methods
A free‐breathing golden‐angle stack‐of‐spirals UTE sequence was used to acquire data for 8 min with prototype and commercial 0.55 T MRI scanners. The data was binned into 12 respiratory phases based on superior–inferior navigator readouts. The previously published iterative motion corrected (iMoCo) reconstruction was extended to include concomitant field correction directly in the cost function. The reconstruction was implemented within the Gadgetron framework for inline reconstruction. Data were retrospectively reconstructed to simulate scan times of 2, 4, 6, and 8 min. Image quality was assessed using apparent SNR and image sharpness. The technique was evaluated in healthy volunteers and patients with known lung pathology including coronavirus disease 2019 infection, chronic granulomatous disease, lymphangioleiomyomatosis, and lung nodules.
Results
The technique provided diagnostic‐quality images, and image quality was maintained with a slight loss in SNR for simulated scan times down to 4 min. Parenchymal apparent SNR was 4.33 ± 0.57, 5.96 ± 0.65, 7.36 ± 0.64, and 7.87 ± 0.65 using iCoMoCo with scan times of 2, 4, 6, and 8 min, respectively. Image sharpness at the diaphragm was comparable between iCoMoCo and reference images. Concomitant field corrections visibly improved the sharpness of anatomical structures away from the isocenter. Inline image reconstruction and artifact correction were achieved in
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA.
- Univ. Bordeaux, CNRS, Centre de Résonance Magnétique des Systèmes Biologiques, UMR 5536, Bordeaux, France.
- Siemens Medical Solutions USA, Malvern, Pennsylvania, USA.
ORCID-Profile: Ahsan Javed, Kelvin Chow, Adrienne Campbell-Washburn
Förderung
- Intramural NIH HHS
- Division of Intramural Research
Zitieren
APA
Javed, A., Ramasawmy, R., Ozenne, V., Su, P., Chow, K., & Campbell-Washburn, A. (2024). Increasing the scan‐efficiency of pulmonary imaging at 0. 55 T using iterative concomitant field and motion‐corrected reconstruction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30054
Vancouver
Javed A, Ramasawmy R, Ozenne V, Su P, Chow K, Campbell-Washburn A. Increasing the scan‐efficiency of pulmonary imaging at 0. 55 T using iterative concomitant field and motion‐corrected reconstruction. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30054
BibTeX
@article{Javed2024,
author = {Javed, Ahsan and Ramasawmy, Rajiv and Ozenne, Valéry and Su, Pan and Chow, Kelvin and Campbell-Washburn, Adrienne},
title = {Increasing the scan‐efficiency of pulmonary imaging at 0. 55 T using iterative concomitant field and motion‐corrected reconstruction},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30054}
}
EUVIMED