Phantom Calibration’s Impact on Measuring Radiodensity Changes: Implications for Evaluating Osteosarcoma Response

Michael A McNamara, Patrick M Mazza, Kendra A Hebert, Samuel S Streeter, Paul M Werth, Jessica M Sin, Rongxiao Zhang, Eric R Henderson

Academic Radiology · 2026-08 · 33 (8) · 3443-3448

DOI: 10.1016/j.acra.2026.03.045 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Academic Radiology, 2026-08. Verfasst von Michael A McNamara und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Dartmouth-Hitchcock Medical Center, One Medical Center Drive, Lebanon, New Hampshire (M.A.M., P.M.M., K.A.H. P.M.W…. Hauptthemen laut MEDLINE: Osteosarcoma, Phantoms, Imaging, Tomography, X-Ray Computed, Bone Neoplasms. Die Arbeit nennt 8 Literaturangaben. Gefördert von: Hitchcock Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

RATIONALE AND OBJECTIVES Improve radiodensity measurements through a controlled, cadaveric study with the primary hypothesis that using high-resolution digital radiographs and image calibration with an aluminum step phantom would be more accurate than in the absence of a phantom. Ultimately, the goal is to improve evaluation of osteosarcoma response to chemotherapy prior to resection using radiographs. MATERIALS AND METHODS Five fresh-frozen cadaveric lower extremities were imaged with computed tomography (CT) and radiography (anteroposterior (AP) and lateral) with varying mAs/kVp before and after distal femoral instrumentation with 4 cm2 rods of known radiodensities, simulating tumors of varying density, and a commercial radiographic calibration step phantom. Using Materials Laboratory for Digital Imaging and Communications in Medicine (DICOM) image analysis, grayscale measurements were obtained for: rods in/out of bone, normal bone proximal to instrumentation, air, and all 16 calibration phantom steps. Radiodensity subtraction accuracy was compared for calibration with all phantom steps, +/- bone or air subtraction. Accuracy was compared using Student T-test and one-way analysis of variance with Tukey HSD post-hoc test. RESULTS Mean accuracy for Cohort 1 (no calibration) (.12, standard deviation [SD] 0.11, N = 120) was better than that of Cohorts 2-4 (calibration) (0.39, SD 0.54-0.60, N = 1920) (p < 0.001, t = 5.4689). Lateral knee radiographs, specific imager settings (4 mAs-70 kVP, 8 mAs-70 kVp, and 4 mAs-80 kV), and variations in bone density/BMI improved accuracy (p < .001). CONCLUSION When measuring radiodensity changes using high-quality, digital images with identical settings, image calibration worsened accuracy, likely by introducing an additional source of potential error. Further work is required to determine if calibration will improve accuracy for images obtained with varying technique.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OsteosarcomaPhantoms, ImagingTomography, X-Ray ComputedBone NeoplasmsHumansCalibrationReproducibility of ResultsCadaverSensitivity and SpecificityRadiographic Image Enhancement

Stichwörter der Autor:innen

CadaverChemotherapyOrthopedic OncologyOsteosarcomaPhantom

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Dartmouth-Hitchcock Medical Center, One Medical Center Drive, Lebanon, New Hampshire (M.A.M., P.M.M., K.A.H. P.M.W., E.R.H.). Electronic address: michael.a.mcnamara@hitchcock.org.
  • Department of Orthopaedic Surgery, Dartmouth-Hitchcock Medical Center, One Medical Center Drive, Lebanon, New Hampshire (M.A.M., P.M.M., K.A.H. P.M.W., E.R.H.).
  • Center for Imaging Medicine, Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire (K.A.H., E.R.H.).
  • Department of Radiology, Dartmouth-Hitchcock Medical Center, One Medical Center Drive, Lebanon, New Hampshire (J.M.S.).
  • Radiation Oncology, Dartmouth Cancer Center, One Medical Center Drive, Lebanon, New Hampshire (R.Z.).
  • Department of Orthopaedic Surgery, Dartmouth-Hitchcock Medical Center, One Medical Center Drive, Lebanon, New Hampshire (M.A.M., P.M.M., K.A.H. P.M.W., E.R.H.); Center for Imaging Medicine, Thayer School of Engineering,…

Förderung

  • Hitchcock Foundation

Zitieren

APA

McNamara, M. A., Mazza, P. M., Hebert, K. A., Streeter, S. S., Werth, P. M., Sin, J. M., Zhang, R., & Henderson, E. R. (2026). Phantom Calibration’s Impact on Measuring Radiodensity Changes: Implications for Evaluating Osteosarcoma Response. Academic Radiology, 33(8), 3443-3448. https://doi.org/10.1016/j.acra.2026.03.045

Vancouver

McNamara MA, Mazza PM, Hebert KA, Streeter SS, Werth PM, Sin JM, et al. Phantom Calibration’s Impact on Measuring Radiodensity Changes: Implications for Evaluating Osteosarcoma Response. Academic Radiology. 2026;33(8):3443-3448. doi:10.1016/j.acra.2026.03.045

BibTeX
@article{McNamara2026,
  author = {McNamara, Michael A and Mazza, Patrick M and Hebert, Kendra A and Streeter, Samuel S and Werth, Paul M and Sin, Jessica M and Zhang, Rongxiao and Henderson, Eric R},
  title = {Phantom Calibration’s Impact on Measuring Radiodensity Changes: Implications for Evaluating Osteosarcoma Response},
  journal = {Academic Radiology},
  year = {2026},
  volume = {33},
  number = {8},
  pages = {3443--3448},
  doi = {10.1016/j.acra.2026.03.045}
}