Material decomposition maps based calibration of dual energy CT scanners for proton therapy planning: a phantom study

David Viar-Hernández iD, Juan Antonio Vera-Sánchez iD, Lucia Schmidt-Santiago iD, Borja Rodriguez-Vila iD, Isabel Lorenzo-Villanueva iD, Elisabet Canals-de-Las-Casas, Juan Castro-Novais iD, Juan Maria Perez-Moreno iD, Fernando Cerrón-Campoo, Norberto Malpica iD, Angel Torrado-Carvajal iD, Alejandro Mazal iD

Physics in Medicine & Biology · 2024

DOI: 10.1088/1361-6560/ad2015 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2024. Verfasst von David Viar-Hernández und 11 weiteren Autor:innen. Beteiligte Einrichtung: Universidad Rey Juan Carlos, Medical Image Analysis and Biometry Laboratory, Madrid, Spain. Hauptthemen laut MEDLINE: Proton Therapy. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: General Electric Precission Healthcare LLC, Spanish Ministry of Science and Innovation, Young Researchers R&D IMPULSO Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We introduce a new calibration method for dual energy CT (DECT) based on material decomposition (MD) maps, specifically iodine and water MD maps. The aim of this method is to provide the first DECT calibration based on MD maps. The experiments were carried out using a general electric (GE) revolution CT scanner with ultra-fast kV switching and used a density phantom by GAMMEX for calibration and evaluation. The calibration process involves several steps. First, we tested the ability of MD values to reproduce Hounsfield unit (HU) values of single energy CT (SECT) acquisitions and it was found that the errors were below 1%, validating their use for HU reproduction. Next, the different definitions of computed Z values were compared and the robustness of the approach based on the materials’ composition was confirmed. Finally, the calibration method was compared with a previous method by Bourque et al , providing a similar level of accuracy and superior performance in terms of precision. Overall, this novel DECT calibration method offers improved accuracy and reliability in determining tissue-specific physical properties. The resulting maps can be valuable for proton therapy treatments, where precise dose calculations and accurate tissue differentiation are crucial for optimal treatment planning and delivery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyTomography, X-Ray ComputedCalibrationReproducibility of ResultsTomography Scanners, X-Ray ComputedPhantoms, Imaging

Stichwörter der Autor:innen

CT calibrationdual energy CTproton therapy

Autor:innen und Institutionen

  • Universidad Rey Juan Carlos, Medical Image Analysis and Biometry Laboratory, Madrid, Spain.
  • Centro de Protonterapia Quironsalud, Servicio de Física Médica, Madrid, Spain.

ORCID-Profile: David Viar-Hernández, Juan Antonio Vera-Sánchez, Lucia Schmidt-Santiago, Borja Rodriguez-Vila, Isabel Lorenzo-Villanueva, Juan Castro-Novais, Juan Maria Perez-Moreno, Norberto Malpica, Angel Torrado-Carvajal, Alejandro Mazal

Förderung

  • General Electric Precission Healthcare LLC
  • Spanish Ministry of Science and Innovation
  • Young Researchers R&D IMPULSO Project
  • Community of Madrid and Rey Juan Carlos University
  • Ministerio de Ciencia e Innovación

Zitieren

APA

Viar-Hernández, D., Vera-Sánchez, J. A., Schmidt-Santiago, L., Rodriguez-Vila, B., Lorenzo-Villanueva, I., Canals-de-Las-Casas, E., Castro-Novais, J., Maria Perez-Moreno, J., Cerrón-Campoo, F., Malpica, N., Torrado-Carvajal, A., & Mazal, A. (2024). Material decomposition maps based calibration of dual energy CT scanners for proton therapy planning: a phantom study. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ad2015

Vancouver

Viar-Hernández D, Vera-Sánchez JA, Schmidt-Santiago L, Rodriguez-Vila B, Lorenzo-Villanueva I, Canals-de-Las-Casas E, et al. Material decomposition maps based calibration of dual energy CT scanners for proton therapy planning: a phantom study. Physics in Medicine & Biology. 2024. doi:10.1088/1361-6560/ad2015

BibTeX
@article{ViarHernndez2024,
  author = {Viar-Hernández, David and Vera-Sánchez, Juan Antonio and Schmidt-Santiago, Lucia and Rodriguez-Vila, Borja and Lorenzo-Villanueva, Isabel and Canals-de-Las-Casas, Elisabet and Castro-Novais, Juan and Maria Perez-Moreno, Juan and Cerrón-Campoo, Fernando and Malpica, Norberto and Torrado-Carvajal, Angel and Mazal, Alejandro},
  title = {Material decomposition maps based calibration of dual energy CT scanners for proton therapy planning: a phantom study},
  journal = {Physics in Medicine & Biology},
  year = {2024},
  doi = {10.1088/1361-6560/ad2015}
}
Kostenlos starten →