EUVIMEDEUVIMED entdecken
Crossref

Ultrahigh-resolution computed tomography of the cervical spine without dose penalty employing a cadmium-telluride photon-counting detector

Nora Conrads, Jan-Peter Grunz, Henner Huflage, Karsten Sebastian Luetkens, Philipp Feldle, Lenhard Pennig, Süleyman Ergün, Thorsten Alexander Bley, Bernhard Petritsch, Andreas Steven Kunz

European Journal of Radiology · 2023

DOI: 10.1016/j.ejrad.2023.110718 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Journal of Radiology, 2023. Verfasst von Nora Conrads und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080 Würzburg, Germany. Electronic add…. Hauptthemen laut MEDLINE: Cadmium, Tomography, X-Ray Computed. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: This cadaveric study compared image quality between a third-generation dual-source CT scanner with energy-integrating detector technology (EID) and a first-generation CT system employing a photon-counting detector (PCD) for the cervical spine in ultrahigh-resolution mode. METHODS: The cervical spine of eight formalin-fixed full-body cadaveric specimens was scanned with both CT systems using 140 kVp scan protocols matched for CTDIvol (full-dose; low-dose; ultralow-dose; 10 mGy; 3 mGy; 1 mGy). Images were reconstructed with 1 mm slice thickness and 0.5 mm increment. Three radiologists rated overall subjective image quality based on an equidistant five-point scale with the intraclass correlation coefficient (ICC) calculated for assessment of interobserver reliability. Contrast-to-noise ratios were calculated individually for bone (CNRbone) and muscle tissue (CNRmuscle) to provide objective criteria of image analysis. RESULTS: Subjective image quality, as well as CNRbone, and CNRmuscle were each superior for PCD-CT compared to EID-CT among dose-matched scan protocol pairs (all p 0.99). Interobserver reliability was good, denoted by an ICC of 0.861 (95 % confidence interval: 0.788 - 0.914; p < 0.001). CONCLUSIONS: In cervical spine examinations, both subjective and objective image quality of PCD-CT were superior to EID-CT in comparison of scan protocols with corresponding dose levels, suggesting potential for significantly reducing the radiation exposure without compromising image quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CadmiumTomography, X-Ray ComputedHumansReproducibility of ResultsPhantoms, ImagingCervical VertebraeCadaver

Stichwörter der Autor:innen

Cervical spineDose reductionPhoton-counting CTSpatial resolution

Substanzen

Cadmium

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080 Würzburg, Germany. Electronic address: Conrads_N1@ukw.de.
  • Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080 Würzburg, Germany.
  • Institute for Diagnostic and Interventional Radiology, Faculty of Medicine and University Hospital Cologne, University of Cologne, Kerpener Straße 62, 50937 Cologne, Germany.
  • Institute of Anatomy and Cell Biology, University of Würzburg, Koellikerstraße 6, 97070 Würzburg, Germany.

Zitieren

APA

Conrads, N., Grunz, J. P., Huflage, H., Luetkens, K. S., Feldle, P., Pennig, L., Ergün, S., Alexander Bley, T., Petritsch, B., & Kunz, A. S. (2023). Ultrahigh-resolution computed tomography of the cervical spine without dose penalty employing a cadmium-telluride photon-counting detector. European Journal of Radiology. https://doi.org/10.1016/j.ejrad.2023.110718

Vancouver

Conrads N, Grunz JP, Huflage H, Luetkens KS, Feldle P, Pennig L, et al. Ultrahigh-resolution computed tomography of the cervical spine without dose penalty employing a cadmium-telluride photon-counting detector. European Journal of Radiology. 2023. doi:10.1016/j.ejrad.2023.110718