A Novel and Reliable Pixel Response Correction Method (DAC-Shifting) for Spectral Photon-Counting CT Imaging
Tomography · 2024
DOI: 10.3390/tomography10070089 ↗Kurz gefasst
Fachartikel in der Zeitschrift Tomography, 2024. Verfasst von Navrit Johan Singh Bal und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Medicine, Aarhus University, 8200 Aarhus, Denmark. Hauptthemen laut MEDLINE: Phantoms, Imaging, Photons, Artifacts. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Novo Nordisk Foundation, Danish Cancer Society, Aarhus University Research Foundation.
Abstract (Originalsprache)
Spectral photon-counting cone-beam computed tomography (CT) imaging is challenged by individual pixel response behaviours, which lead to noisy projection images and subsequent image artefacts like rings. Existing methods to correct for this either use calibration measurements, like signal-to-thickness calibration (STC), or perform a post-processing ring artefact correction of sinogram data or scan reconstructions without taking the pixel response explicitly into account. Here, we present a novel post-processing method (digital-to-analogue converter (DAC)-shifting) which explicitly measures the current pixel response using flat-field images and subsequently corrects the projection data. The DAC-shifting method was evaluated using a repeat series of the spectral photon-counting imaging (Medipix3) of a phantom with different density inserts and iodine K-edge imaging. The method was also compared against polymethyl methacrylate (PMMA)-based STC. The DAC-shifting method was shown to be effective in correcting individual pixel responses and was robust against detector instability; it led to a 47.4% average reduction in CT-number variation in homogeneous materials, with a range of 40.7–55.6%. On the contrary, the STC correction showed varying results; a 13.7% average reduction in CT-number variation, ranging from a 43.7% increase to a 45.5% reduction. In K-edge imaging, DAC-shifting provides a sharper attenuation peak and more uniform CT values, which are expected to benefit iodine concentration quantifications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Clinical Medicine, Aarhus University, 8200 Aarhus, Denmark.
- Danish Centre for Particle Therapy, Aarhus University Hospital, 8200 Aarhus, Denmark.
- Department of Pathology, Aarhus University Hospital, 8200 Aarhus, Denmark.
ORCID-Profile: Navrit Johan Singh Bal, Imaiyan Chitra Ragupathy, Trine Tramm, Jasper Nijkamp
Förderung
- Novo Nordisk Foundation
- Danish Cancer Society
- Aarhus University Research Foundation
Zitieren
APA
Bal, N. J. S., Chitra Ragupathy, I., Tramm, T., & Nijkamp, J. (2024). A Novel and Reliable Pixel Response Correction Method (DAC-Shifting) for Spectral Photon-Counting CT Imaging. Tomography. https://doi.org/10.3390/tomography10070089
Vancouver
Bal NJS, Chitra Ragupathy I, Tramm T, Nijkamp J. A Novel and Reliable Pixel Response Correction Method (DAC-Shifting) for Spectral Photon-Counting CT Imaging. Tomography. 2024. doi:10.3390/tomography10070089
BibTeX
@article{Bal2024,
author = {Bal, Navrit Johan Singh and Chitra Ragupathy, Imaiyan and Tramm, Trine and Nijkamp, Jasper},
title = {A Novel and Reliable Pixel Response Correction Method (DAC-Shifting) for Spectral Photon-Counting CT Imaging},
journal = {Tomography},
year = {2024},
doi = {10.3390/tomography10070089}
}
EUVIMED