EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Improved commissioning of lung stereotactic body radiotherapy using a customized respiratory motion Phantom: a single- institutional study

Ashlesha Gill, Mohamed Nawar, Pejman Rowshanfarzad, Andrew Hirst, Malgorzata Skorska, Tom Milan, Nicholas Bucknell, Mahsheed Sabet

Physical and Engineering Sciences in Medicine · 2025

DOI: 10.1007/s13246-025-01550-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2025. Verfasst von Ashlesha Gill und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Physics, Mathematics and Computing, The University of Western Australia, 35 Stirling Highway Crawley, Crawley, WA, 6009, Australia. ashlesh…. Hauptthemen laut MEDLINE: Radiosurgery, Phantoms, Imaging, Respiration, Lung. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Western Australia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Stereotactic body radiation therapy (SBRT) involves delivering high doses of radiation with geometric precision in a few hypofractionated schedules. In lung SBRT, respiratory motion is an additional concern as it could cause the delivered dose distribution to deviate from the treatment plan. Therefore, it is crucial to conduct accurate commissioning tests on a dynamic phantom. In this study, the QUASAR™ Respiratory Motion Phantom was customized using 3D-printed parts to minimize motion-induced errors in measurements. The customisations included a specialized ion chamber insert designed to move with the tumour and measure the average dose at its centre. A film insert was also developed for secure fixation, enabling precise dose verification on a static plane while minimizing the risk of friction-related damage. The quality assurance (QA) tests were performed on the plans created for phantom studies indicated that ion chamber measurements were within 1.9% of the planned dose, and film gamma analysis demonstrated pass rates over 95% using the 3%/1 mm criteria. A set of SBRT volumetric modulated arc therapy (VMAT) plans were created for a suite of test patients using both flattened and flattening filter free (FFF) 6 MV beams and utilising robust optimization. A standardized patient-specific QA protocol was used to evaluate the treatment plans of 20 test patients, yielding film gamma pass rates above 98.8%. The suggested approach, using the 3D-printed inserts, effectively mitigated dose-blurring, providing a robust tool for lung SBRT commissioning and ensuring the reliability of lung cancer treatment with SBRT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryPhantoms, ImagingRespirationLungHumansLung NeoplasmsRadiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

3D printingCommissioningDose blurringLungSBRTTumour motion

Autor:innen und Institutionen

  • School of Physics, Mathematics and Computing, The University of Western Australia, 35 Stirling Highway Crawley, Crawley, WA, 6009, Australia. ashlesha.gill@research.uwa.edu.au.
  • Centre for Advanced Technologies in Cancer Research (CATCR), Perth, WA, Australia. ashlesha.gill@research.uwa.edu.au.
  • Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, Australia.
  • School of Physics, Mathematics and Computing, The University of Western Australia, 35 Stirling Highway Crawley, Crawley, WA, 6009, Australia.
  • Centre for Advanced Technologies in Cancer Research (CATCR), Perth, WA, Australia.

Förderung

  • University of Western Australia

Zitieren

APA

Gill, A., Nawar, M., Rowshanfarzad, P., Hirst, A., Skorska, M., Milan, T., Bucknell, N., & Sabet, M. (2025). Improved commissioning of lung stereotactic body radiotherapy using a customized respiratory motion Phantom: a single- institutional study. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-025-01550-0

Vancouver

Gill A, Nawar M, Rowshanfarzad P, Hirst A, Skorska M, Milan T, et al. Improved commissioning of lung stereotactic body radiotherapy using a customized respiratory motion Phantom: a single- institutional study. Physical and Engineering Sciences in Medicine. 2025. doi:10.1007/s13246-025-01550-0

BibTeX
@article{Gill2025,
  author = {Gill, Ashlesha and Nawar, Mohamed and Rowshanfarzad, Pejman and Hirst, Andrew and Skorska, Malgorzata and Milan, Tom and Bucknell, Nicholas and Sabet, Mahsheed},
  title = {Improved commissioning of lung stereotactic body radiotherapy using a customized respiratory motion Phantom: a single- institutional study},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2025},
  doi = {10.1007/s13246-025-01550-0}
}
Kostenlos starten →