Effects of cardiac motion on dose distribution during stereotactic arrhythmia radioablation treatment: A simulation and phantom study
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70021 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Takayuki Miyachi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan. Hauptthemen laut MEDLINE: Radiosurgery, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Arrhythmias, Cardiac. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Cardiac motion may degrade dose distribution during stereotactic arrhythmia radioablation using the CyberKnife system, a robotic radiosurgery system. This study evaluated the dose distribution changes using a self‐made cardiac dynamic platform that mimics cardiac motion.
Methods
The cardiac dynamic platform was operated with amplitudes of 5 and 3.5 mm along the superior–inferior (SI) and left–right (LR) directions, respectively. The respiratory motion tracking of the CyberKnife system was applied when respiratory motion, simulated using a commercial platform, was introduced. The accuracy of respiratory motion tracking was evaluated by the correlation error between infrared markers and a fiducial marker. The dose distribution was compared with and without cardiac motion. The evaluations included error in the centroid analysis of the irradiated dose distribution, dose profile analysis in the SI and LR directions, and dose distribution analysis comparing the irradiated and planned dose distributions.
Results
Cardiac motion increased the correlation error in the direction of motion. Cardiac motion displaced the centroid by up to 0.23 and 0.19 mm in the SI and LR directions, respectively. Cardiac motion blurring caused the distance of the isodose lines to become smaller (bigger) at higher (lower) doses in the SI direction. The gamma pass rate was reduced by cardiac motion but exceeded 94.1% with 1 mm/3% for all conditions. Respiratory motion tracking was also effective under cardiac motion. The cardiac motion slightly varied the dose at the edges of the irradiation volume.
Conclusion
While cardiac motion increased respiratory tracking correlation errors, its effects on dose distribution were limited in this study. Further studies using motion phantoms that are close to a human or individual patient are necessary for a more detailed understanding of the effects of cardiac motion.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan.
- Radioisotope Research Center, Nagoya University, Nagoya, Aichi, Japan.
- Department of Radiological Technology, Nagoya University Hospital, Nagoya, Aichi, Japan.
Zitieren
APA
Miyachi, T., Kamomae, T., Kawabata, F., Okudaira, K., Kawamura, M., Ishihara, S., & Naganawa, S. (2025). Effects of cardiac motion on dose distribution during stereotactic arrhythmia radioablation treatment: A simulation and phantom study. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70021
Vancouver
Miyachi T, Kamomae T, Kawabata F, Okudaira K, Kawamura M, Ishihara S, et al. Effects of cardiac motion on dose distribution during stereotactic arrhythmia radioablation treatment: A simulation and phantom study. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70021
BibTeX
@article{Miyachi2025,
author = {Miyachi, Takayuki and Kamomae, Takeshi and Kawabata, Fumitaka and Okudaira, Kuniyasu and Kawamura, Mariko and Ishihara, Shunichi and Naganawa, Shinji},
title = {Effects of cardiac motion on dose distribution during stereotactic arrhythmia radioablation treatment: A simulation and phantom study},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70021}
}
EUVIMED