Kurz gefasst

Evaluationsstudie in der Zeitschrift Physical and Engineering Sciences in Medicine, 2024. Verfasst von O B Kolcu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Istinye University, TR-34010, Istanbul, Turkey. onur.kolcu@istinye.edu.tr. Hauptthemen laut MEDLINE: Scintillation Counting, Gamma Rays, Monte Carlo Method. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: Türkiye Bilimsel ve Teknolojik Arastirma Kurumu.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The gamma probe is a commonly used detector for localizing sentinel lymph nodes after the injection of radiopharmaceuticals. In recent years, studies have focused on improving the features of gamma probes to achieve more consistent localization of the radiotracer uptake. As part of this effort, a novel gamma probe prototype based on an active shielding was developed, and its characteristics, including sensitivity, resolution and shielding effectiveness, were determined. The prototype integrates trapezoidal-shaped bismuth germanate (BGO) array coupled with a silicon photomultiplier (SiPM) array, accompanied by dedicated electronics and software for stand alone usage. We conducted a thorough characterization, validating experimental observations through Monte Carlo simulations using the GEANT4 simulation package. In scattering environment, with a probe-source distance of 30 mm, the experimental results show that the detector sensitivity is 120 ± 5 cps/MBq, and the spatial and angular resolutions, in terms of full width at half maximum (FWHM), are 44.8 ± 1.3 mm and 87.3 ± 1 . 5 ∘ , respectively. The shielding effectiveness of the probe was determined to be greater than 95 % . The prototype with active shielding was found to have comparable performance to conventional gamma probes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Scintillation CountingGamma RaysMonte Carlo MethodRadiation ProtectionEquipment Design

Stichwörter der Autor:innen

Active shieldingGamma probesIntra-operative probesSentinel lymph node

Autor:innen und Institutionen

  • Istinye University, TR-34010, Istanbul, Turkey. onur.kolcu@istinye.edu.tr.
  • Yildiz Technical University, TR-34220, Istanbul, Turkey.
  • Istanbul Technical University, TR-34467, Istanbul, Turkey.
  • Mimar Sinan Fine Arts University, TR-34380, Istanbul, Turkey.
  • Istinye University, TR-34010, Istanbul, Turkey.

Förderung

  • Türkiye Bilimsel ve Teknolojik Arastirma Kurumu

Zitieren

APA

Kolcu, O. B., Yetkin, T., Zengin, A. T., Iren, E., & Günay, E. C. (2024). Development and performance evaluation of a novel scintillation-based active shielding gamma probe. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-024-01474-1

Vancouver

Kolcu OB, Yetkin T, Zengin AT, Iren E, Günay EC. Development and performance evaluation of a novel scintillation-based active shielding gamma probe. Physical and Engineering Sciences in Medicine. 2024. doi:10.1007/s13246-024-01474-1

BibTeX
@article{Kolcu2024,
  author = {Kolcu, O B and Yetkin, T and Zengin, A T and Iren, E and Günay, E C},
  title = {Development and performance evaluation of a novel scintillation-based active shielding gamma probe},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2024},
  doi = {10.1007/s13246-024-01474-1}
}
Kostenlos starten →