EUVIMEDEUVIMED entdecken
Crossref

First-in-man clinical implantation and acute pacing performance of a novel subxiphoidal pacemaker system

Imad Libbus, Venkatakrishna Tholakanahalli, Henri Roukoz, Varun Verma, Divya Isac, Yatheendhar Manicka

Journal of Interventional Cardiac Electrophysiology · 2024

DOI: 10.1007/s10840-024-01815-w ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Interventional Cardiac Electrophysiology, 2024. Verfasst von Imad Libbus und 5 weiteren Autor:innen. Beteiligte Einrichtung: Calyan Technologies, Saint Paul, MN, USA. imad@calyantech.com. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Cardiac Pacing, Artificial, Bradycardia, Prosthesis Implantation. Die Arbeit nennt 8 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Intravascular cardiac pacemakers are an established therapy for bradycardic indications. Recently, a new class of leadless pacemakers have mitigated some of the complications related to pacemaker leads. In this study, we evaluate the implantation and performance of a novel extravascular technology that delivers epicardial pacing through a subxiphoidal approach. METHODS: Fifteen patients undergoing non-emergent open-heart surgery were enrolled. A midline incision was made just below the xiphoid process, and substernal tunneling was used to create a pocket for the device and to access the anterior pericardium over the right ventricle. The test device (Calyan Technologies, Oakdale, MN) was temporarily inserted underneath the ribcage and clipped to the xiphoid process. The sensing and pacing electrode (FlexArm) was positioned on the anterior pericardium. Ventricular sensing and pacing capture thresholds were measured. RESULTS: The test device was successfully implanted in all 15 patients. There were no device or procedure-related adverse events. The first five implanted patients had no pacing capture at maximum stimulation intensity. Design changes were made to the device, including different electrode size and shape, and successful ventricular capture was achieved in 9 of the subsequent 10 patients. In these patients, pacing threshold was 3.8 ± 1.6 mA with a pulse width of 0.5 ms. All devices were successfully explanted at the end of the procedure. CONCLUSIONS: In a first-in-human experience with a novel extravascular pacemaker, this study demonstrated the feasibility of pericardial ventricular pacing via a subxiphoidal approach. Further chronic studies are required to evaluate the safety and performance of this novel pacing technology.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialCardiac Pacing, ArtificialBradycardiaProsthesis ImplantationHumansFemaleMaleTreatment OutcomeAgedEquipment DesignMiddle AgedAged, 80 and overEquipment Failure Analysis

Stichwörter der Autor:innen

ArrhythmiasBradycardiaCardiac pacingLeadless pacingPacemaker

Autor:innen und Institutionen

  • Calyan Technologies, Saint Paul, MN, USA. imad@calyantech.com.
  • Calyan Technologies, 7300 Hudson Boulevard North, Suite 290, Oakdale, MN, 55128, USA. imad@calyantech.com.
  • University of Minnesota, Minneapolis, MN, USA.
  • Calyan Technologies, Saint Paul, MN, USA.

Zitieren

APA

Libbus, I., Tholakanahalli, V., Roukoz, H., Verma, V., Isac, D., & Manicka, Y. (2024). First-in-man clinical implantation and acute pacing performance of a novel subxiphoidal pacemaker system. Journal of Interventional Cardiac Electrophysiology. https://doi.org/10.1007/s10840-024-01815-w

Vancouver

Libbus I, Tholakanahalli V, Roukoz H, Verma V, Isac D, Manicka Y. First-in-man clinical implantation and acute pacing performance of a novel subxiphoidal pacemaker system. Journal of Interventional Cardiac Electrophysiology. 2024. doi:10.1007/s10840-024-01815-w

BibTeX
@article{Libbus2024,
  author = {Libbus, Imad and Tholakanahalli, Venkatakrishna and Roukoz, Henri and Verma, Varun and Isac, Divya and Manicka, Yatheendhar},
  title = {First-in-man clinical implantation and acute pacing performance of a novel subxiphoidal pacemaker system},
  journal = {Journal of Interventional Cardiac Electrophysiology},
  year = {2024},
  doi = {10.1007/s10840-024-01815-w}
}