RF phase modulation improves quantitative transient state sequences under constrained conditions
Magnetic Resonance Materials in Physics, Biology and Medicine · 2025
DOI: 10.1007/s10334-025-01293-9 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2025. Verfasst von Miha Fuderer und 4 weiteren Autor:innen. Beteiligte Einrichtung: Computational Imaging Group for MR Diagnostics & Therapy, Center for Image Sciences, University Medical Center Utrecht, Heidelberglaan 100, 3585CX, U…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging. Die Arbeit nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek.
Abstract (Originalsprache)
Objective
Within gradient-spoiled transient-state MR sequences like Magnetic Resonance Fingerprinting or Magnetic Resonance Spin TomogrAphy in Time-domain (MR-STAT), it is examined whether an optimized RF phase modulation can help to improve the precision of the resulting relaxometry maps.
Methods
Using a Cramer-Rao based method called BLAKJac, optimized sequences of RF pulses have been generated for two scenarios (amplitude-only modulation and amplitude + phase modulation) and for several conditions. These sequences have been tested on a phantom, a healthy human brain and a healthy human leg, to reconstruct parametric maps ( $${T}_{1}$$ T 1 and $${T}_{2}$$ T 2 ) as well as their standard deviations.
Results
The amplitude + phase modulation scenario systematically resulted in lower noise levels than the amplitude-only modulation scenario. On average, the difference was around 34%, but it was substantially larger for scans acquired under SAR restrictions. Compared to amplitude-only, in the amplitude + phase modulation scenario, the relevance of an inversion pulse and of a pause were greatly reduced, at least considering overall precision and in-phantom accuracy.
Conclusion
The application of an optimized RF phase modulation in quantitative transient-states MRI is beneficial for almost all tested scenarios and conditions, in particular under SAR restrictions Furthermore, RF phase modulation reduces the need for inversions pulses and pauses.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Computational Imaging Group for MR Diagnostics & Therapy, Center for Image Sciences, University Medical Center Utrecht, Heidelberglaan 100, 3585CX, Utrecht, The Netherlands. m.fuderer@umcutrecht.nl.
- Computational Imaging Group for MR Diagnostics & Therapy, Center for Image Sciences, University Medical Center Utrecht, Heidelberglaan 100, 3585CX, Utrecht, The Netherlands.
Förderung
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
Zitieren
APA
Fuderer, M., Liu, H., van der Heide, O., van den Berg, C. A. T., & Sbrizzi, A. (2025). RF phase modulation improves quantitative transient state sequences under constrained conditions. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-025-01293-9
Vancouver
Fuderer M, Liu H, van der Heide O, van den Berg CAT, Sbrizzi A. RF phase modulation improves quantitative transient state sequences under constrained conditions. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025. doi:10.1007/s10334-025-01293-9
BibTeX
@article{Fuderer2025,
author = {Fuderer, Miha and Liu, Hongyan and van der Heide, Oscar and van den Berg, Cornelis A T and Sbrizzi, Alessandro},
title = {RF phase modulation improves quantitative transient state sequences under constrained conditions},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2025},
doi = {10.1007/s10334-025-01293-9}
}
EUVIMED