Diffusion Weighted Fast Spin Echo With RF ‐Encoded Slabs for Portable Low‐Field MRI Systems
Magnetic Resonance in Medicine · 2026-06 · 95 (6) · 3253-3269
DOI: 10.1002/mrm.70292 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 25.02.2026. Verfasst von Philip K Lee und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, People's Republic of China. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Brain. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 73 Literaturangaben. Gefördert von: National Key R&D Program of China, National Natural Science Foundation of China, Shanghai Jiao Tong University.
Abstract (Originalsprache)
Purpose
To develop a robust diffusion weighted (DW) acquisition with clinically feasible scan times for portable MRI systems.
Methods
A linear reconstruction model including: tailored RF‐encoding schemes, motion‐induced phases, and systematic phase errors was developed for slab diffusion weighted fast spin echo (FSE) data acquired with quadratic phase increment. Systematic phases imparted by diffusion gradients were obtained using a calibration phantom. Trace‐weighted DW‐FSE was compared to DW‐EPI in vivo on a portable 110 mT H‐based system. A high‐NEX DW‐FSE acquisition was performed in healthy volunteers followed by Monte Carlo retrospective undersampling to evaluate expected image quality and mean diffusivity variation in a DWI protocol with clinically feasible scan times.
Results
CPMG calibration and polynomials fitted to systematic phase errors enabled accurate estimation of motion‐induced phases. Accounting of different phase sources reduced ghosting artifacts and improved SNR efficiency by enabling RF‐encoding. RF‐encoded slab DW‐FSE obtained in vivo trace‐weighted images without distortion artifacts and with voxel sizes 5× smaller than DW‐EPI. The retrospectively undersampled acquisition showed that 12 cm slice coverage with voxel size 2.5 × 2.5 × 5 mm 3 could be achieved in 12 min scan time. The mean diffusivity standard deviation between Monte Carlo trials was 0.2 × 10 −3 mm 2 /s.
Conclusion
Diffusion weighted FSE is a robust, distortionless approach for high SNR efficiency DWI on portable MRI systems. Multiband quadratic phase increment DW‐FSE leverages the low SAR and higher peak B 1 provided by low field strengths and is less dependent on performant hardware compared to DW‐EPI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, People's Republic of China.
ORCID-Profile: Philip K Lee, Yueqi Qiu, Zhiyong Zhang
Förderung
- National Key R&D Program of China
- National Natural Science Foundation of China
- Shanghai Jiao Tong University
- National Key Research and Development Program of China
Zitieren
APA
Lee, P. K., Qiu, Y., Chen, S., Wang, C., & Zhang, Z. (2026). Diffusion Weighted Fast Spin Echo With RF ‐Encoded Slabs for Portable Low‐Field MRI Systems. Magnetic Resonance in Medicine, 95(6), 3253-3269. https://doi.org/10.1002/mrm.70292
Vancouver
Lee PK, Qiu Y, Chen S, Wang C, Zhang Z. Diffusion Weighted Fast Spin Echo With RF ‐Encoded Slabs for Portable Low‐Field MRI Systems. Magnetic Resonance in Medicine. 2026;95(6):3253-3269. doi:10.1002/mrm.70292
BibTeX
@article{Lee2026,
author = {Lee, Philip K and Qiu, Yueqi and Chen, Suen and Wang, Changyue and Zhang, Zhiyong},
title = {Diffusion Weighted Fast Spin Echo With RF ‐Encoded Slabs for Portable Low‐Field MRI Systems},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {95},
number = {6},
pages = {3253--3269},
doi = {10.1002/mrm.70292}
}