RF coil design strategies for improving SNR at the ultrahigh magnetic field of 10.5T
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30315 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Matt Waks und 18 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Signal-To-Noise Ratio, Equipment Design. Die Arbeit wurde bisher 31-mal zitiert (Crossref) und nennt 81 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, NCRR NIH HHS, NINDS NIH HHS.
Abstract (Originalsprache)
Purpose
Toward pushing the boundaries of ultrahigh fields for human brain imaging, we wish to evaluate experimentally achievable SNR relative to ultimate intrinsic SNR (uiSNR) at 10.5T, develop design strategies toward approaching the latter, quantify magnetic field–dependent SNR gains, and demonstrate the feasibility of whole‐brain, high‐resolution human brain imaging at this uniquely high field strength.
Methods
A dual row 16‐channel self‐decoupled transmit (Tx) and receive (Rx) array was developed for 10.5T using custom Tx/Rx switches. A 64‐channel receive‐only array was built to fit into the 16‐channel Tx/Rx array. Electromagnetic modeling and experiments were used to define safe operational power limits. Experimental SNR was evaluated relative to uiSNR at 10.5T and 7T.
Results
The 64‐channel Rx array alone captured approximately 50% of the central uiSNR at 10.5T, while an identical array developed for 7T captured about 76% of uiSNR at 7T. The 16‐channel Tx/80‐channel Rx configuration brought the fraction of uiSNR captured at 10.5T to levels comparable to the 64‐channel Rx array at 7T. SNR data displayed an approximate dependence over a large central region when evaluated in the context of uiSNR. Whole‐brain, high‐resolution ‐weighted and T 1 ‐weighted anatomical and gradient‐recalled‐echo BOLD‐EPI functional MRI images were obtained at 10.5T for the first time with such an advanced array.
Conclusion
We demonstrated the ability to approach the uiSNR at 10.5T over the human brain, achieving large SNR gains over 7T, currently the most commonly used ultrahigh‐field platform. Whole‐brain, high‐resolution anatomical and EPI‐based functional MRI data were obtained at 10.5T, illustrating the promise of greater than 10T fields in studying the human brain.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA.
- Center for Advanced Imaging Innovation and Research (CAI2R) and Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
ORCID-Profile: Matt Waks, Russell L Lagore, Edward Auerbach, Andrea Grant, Alireza Sadeghi-Tarakameh, Lance DelaBarre, Steve Jungst, Nader Tavaf, Riccardo Lattanzi, Ilias Giannakopoulos, Steen Moeller, Xiaoping Wu, Essa Yacoub, Luca Vizioli, Simon Schmidt, Gregory J Metzger, Yigitcan Eryaman, Gregor Adriany, Kamil Uğurbil
Förderung
- NIBIB NIH HHS
- NCRR NIH HHS
- NINDS NIH HHS
- National Institutes of Health
- National Institute of Neurological Disorders and Stroke
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Waks, M., Lagore, R. L., Auerbach, E., Grant, A., Sadeghi-Tarakameh, A., DelaBarre, L., Jungst, S., Tavaf, N., Lattanzi, R., Giannakopoulos, I., Moeller, S., Wu, X., Yacoub, E., Vizioli, L., Schmidt, S., Metzger, G. J., Eryaman, Y., Adriany, G., & Uğurbil, K. (2025). RF coil design strategies for improving SNR at the ultrahigh magnetic field of 10.5T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30315
Vancouver
Waks M, Lagore RL, Auerbach E, Grant A, Sadeghi-Tarakameh A, DelaBarre L, et al. RF coil design strategies for improving SNR at the ultrahigh magnetic field of 10.5T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30315
BibTeX
@article{Waks2025,
author = {Waks, Matt and Lagore, Russell L and Auerbach, Edward and Grant, Andrea and Sadeghi-Tarakameh, Alireza and DelaBarre, Lance and Jungst, Steve and Tavaf, Nader and Lattanzi, Riccardo and Giannakopoulos, Ilias and Moeller, Steen and Wu, Xiaoping and Yacoub, Essa and Vizioli, Luca and Schmidt, Simon and Metzger, Gregory J and Eryaman, Yigitcan and Adriany, Gregor and Uğurbil, Kamil},
title = {RF coil design strategies for improving SNR at the ultrahigh magnetic field of 10.5T},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30315}
}