Incorporating Spatial and Spectral Saturation Modules Into MR Fingerprinting
NMR in Biomedicine · 2025
DOI: 10.1002/nbm.70000 ↗Kurz gefasst
Fachartikel in der Zeitschrift NMR in Biomedicine, 2025. Verfasst von Christopher G Trimble und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Circulation and Medical Imaging, Norwegian University of Science and Technology, Trondheim, Norway. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging. Die Arbeit nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Liaison Committee between the Central Norway Regional Health Authority and the Norwegian University of Science and Technology, Australian Research Council, Kreftforeningen.
Abstract (Originalsprache)
In this work, we introduce spatial and chemical saturation options for artefact reduction in magnetic resonance fingerprinting (MRF) and assess their impact on T 1 and T 2 mapping accuracy. An existing radial MRF pulse sequence was modified to enable spatial and chemical saturation. Phantom experiments were performed to demonstrate flow artefact reduction and evaluate the accuracy of the T 1 and T 2 maps. As an in vivo demonstration, MRF of the prostate was performed on an asymptomatic volunteer using saturation modules to reduce flow‐related artefacts. T 1 , T 2 and B 1 + maps obtained with and without saturation modules were compared. Application of spatial saturation in prostate MRF reduced streaking artefacts from the femoral vessels. When saturation is enabled T 1 accuracy is preserved, and T 2 accuracy remains acceptable up to approximately 100 ms. Chemical and spatial saturation can be incorporated into MRF sequences with limited impact on T 1 accuracy. Further sequence optimisation may be needed to accurately estimate long T 2 components. Spatial saturation modules have potential in prostate MRF applications as a means to reduce flow‐related artefacts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Circulation and Medical Imaging, Norwegian University of Science and Technology, Trondheim, Norway.
- Department of Radiology and Nuclear Medicine, St. Olavs Hospital, Trondheim University Hospital, Trondheim, Norway.
- Centre for Advanced Imaging, The University of Queensland, St Lucia, Queensland, Australia.
- ARC Training Centre for Innovation on Biomedical Imaging Technology (CIBIT), The University of Queensland, St Lucia, Queensland, Australia.
- School of Electrical Engineering and Computer Science, The University of Queensland, St Lucia, Queensland, Australia.
- Computational Imaging Group for MR Diagnostics and Therapy, Department of Radiotherapy, UMC Utrecht, Utrecht, The Netherlands.
ORCID-Profile: Christopher G Trimble, Kaia I Sørland, Tone F Bathen
Förderung
- Liaison Committee between the Central Norway Regional Health Authority and the Norwegian University of Science and Technology
- Australian Research Council
- Kreftforeningen
- Prostatakreftforeningen
- Norges Forskningsråd
- Research Council of Norway
Zitieren
APA
Trimble, C. G., Sørland, K. I., Wu, C. Y., Riel, M. H. C. V., Bathen, T. F., Elschot, M., & Cloos, M. A. (2025). Incorporating Spatial and Spectral Saturation Modules Into MR Fingerprinting. NMR in Biomedicine. https://doi.org/10.1002/nbm.70000
Vancouver
Trimble CG, Sørland KI, Wu CY, Riel MHCV, Bathen TF, Elschot M, et al. Incorporating Spatial and Spectral Saturation Modules Into MR Fingerprinting. NMR in Biomedicine. 2025. doi:10.1002/nbm.70000
BibTeX
@article{Trimble2025,
author = {Trimble, Christopher G and Sørland, Kaia I and Wu, Chia-Yin and Riel, Max H C van and Bathen, Tone F and Elschot, Mattijs and Cloos, Martijn A},
title = {Incorporating Spatial and Spectral Saturation Modules Into MR Fingerprinting},
journal = {NMR in Biomedicine},
year = {2025},
doi = {10.1002/nbm.70000}
}
EUVIMED