Comparison of Amplitude‐ and Amplitude‐Frequency‐Modulated Pulses for Outer Volume Suppression in Cardiac MRI : Application to Myocardial T2 Mapping
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70515 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Ayda Arami und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging Physics, Delft University of Technology, Delft, the Netherlands. Hauptthemen laut MEDLINE: Heart, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek, The 4TU Precision Medicine program supported by High Tech for a Sustainable Future, Nederlandse Hartstichting.
Abstract (Originalsprache)
Purpose
To evaluate single‐ and dual‐band pulse modules for suppressing extra‐cardiac signal in cardiac MRI, and test their applicability for time‐efficient imaging with reduced FOV in myocardial mapping at 3 T.
Methods
Outer Volume Suppression (OVS) was implemented using two regional saturation slabs, along the phase‐encoding direction. Different Amplitude‐Modulated (AM) and Amplitude‐Frequency‐Modulated (AM‐FM) pulses were explored. AM sinc and AM‐FM Hyperbolic Secant (HS) pulses, as single‐ (S:1b and HS:1b) and dual‐band (S:2b and HS:2b) versions, were chosen for full evaluation. Pulses were individually optimized for minimal residual signal and optimal suppression‐band profiles. Pulse performance was assessed through simulations and phantom experiments, and evaluated in seven healthy subjects using residual signal and quantitative myocardial mapping with rFOV.
Results
In simulations, optimized HS pulses demonstrated superior suppression‐band profile homogeneity and robustness, while both pulse classes showed comparable sensitivity. Phantom data revealed better signal suppression with HS pulses (S:1b: 19.22% ± 3.23%, S:2b: 16.85% ± 6.80%, HS:1b: 4.66% ± 1.04%, HS:2b: 2.85% ± 1.01%). The residual difference across slabs was slightly reduced using simultaneous dual‐band compared with sequential single‐band approaches (S:1b: 3.81% vs. S:2b: 0.92%; HS:1b: 4.56% vs. HS:2b: 0.66%). In vivo, despite significantly lower residual with HS pulses ( p = 0.0015) and visually observed artifacts with sinc pulses, no significant difference in was observed across all configurations ( p = 0.087) in myocardial mapping.
Conclusion
AM‐FM pulses achieved robust suppression of extra‐cardiac signals, with dual‐band implementations minimizing suppression‐band differences. This can enable artifact‐free rFOV imaging, for improved image quality in myocardial mapping.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Imaging Physics, Delft University of Technology, Delft, the Netherlands.
- Department of Radiology, Leiden University Medical Center, Leiden, the Netherlands.
- Holland PTC, Delft, the Netherlands.
- Philips Benelux, Eindhoven, the Netherlands.
ORCID-Profile: Ayda Arami, Maša Božić-Iven, Yi Zhang, Qian Tao, Hildebrandus Johannes Lamb, Sebastian Weingärtner
Förderung
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
- The 4TU Precision Medicine program supported by High Tech for a Sustainable Future
- Nederlandse Hartstichting
- LUMC-TU Delft Delft Health Initiative
- European Research Council
Zitieren
APA
Arami, A., Božić-Iven, M., van de Steeg-Henzen, C., Zhao, Y., Zhang, Y., Tao, Q., Cernicanu, A., Lamb, H. J., & Weingärtner, S. (2026). Comparison of Amplitude‐ and Amplitude‐Frequency‐Modulated Pulses for Outer Volume Suppression in Cardiac MRI : Application to Myocardial T2 Mapping. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70515
Vancouver
Arami A, Božić-Iven M, van de Steeg-Henzen C, Zhao Y, Zhang Y, Tao Q, et al. Comparison of Amplitude‐ and Amplitude‐Frequency‐Modulated Pulses for Outer Volume Suppression in Cardiac MRI : Application to Myocardial T2 Mapping. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70515
BibTeX
@article{Arami2026,
author = {Arami, Ayda and Božić-Iven, Maša and van de Steeg-Henzen, Christal and Zhao, Yidong and Zhang, Yi and Tao, Qian and Cernicanu, Alexandru and Lamb, Hildebrandus Johannes and Weingärtner, Sebastian},
title = {Comparison of Amplitude‐ and Amplitude‐Frequency‐Modulated Pulses for Outer Volume Suppression in Cardiac MRI : Application to Myocardial T2 Mapping},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70515}
}
EUVIMED