Fat/Water Separation at 7 T Using a 3D Radial Sequence With Quasi‐Continuous Echo Times
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70323 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Matthias Rohe und 9 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Hauptthemen laut MEDLINE: Adipose Tissue, Imaging, Three-Dimensional, Magnetic Resonance Imaging, Body Water. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Interdisciplinary Center for Clinical Research (IZKF), German Research Foundation (DFG), Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Purpose
To achieve a reliable fat/water separation (FWS) at 7 T using a new 3D radial sequence and reconstruction workflow with quasi‐continuous TE sampling.
Methods
A 3D radial density‐adapted sequence with quasi‐continuous TE sampling was developed for a 7 T whole‐body MRI system. The reconstruction workflow included an off‐resonance correction to reduce chemical shift artifacts. Fat and water signals were separated using a graph cut algorithm, yielding proton density fat fraction maps. The sequence and reconstruction pipeline were tested and validated using phantom measurements and in vivo acquisitions of the lower leg of healthy volunteers.
Results
The proposed sequence and reconstruction pipeline enabled sampling of the fat/water oscillation curve over a range from a minimal mean TE of 0.27 ms to a maximum mean TE of 10.13 ms, with an effective TE increment of 85 μs. The off‐resonant correction significantly reduced chemical shift artifacts in the fat signal. The data points of the fat/water oscillation showed good agreement with a multi‐peak fit function. The quantification of phantom fat percentages was verified with a mean absolute error of 1.5%. The proposed pipeline resulted in consistent interpretation of fat and water signals without any swaps, neither in phantom nor in in vivo measurements.
Conclusion
This study successfully demonstrated the application of a 3D radial sequence with quasi‐continuous TEs for FWS at 7 T. Due to the high sampling rate (effective TE increment below 100 μs), the presented FWS workflow demonstrated high robustness against fat/water swaps, which are typical at ultra‐high field strengths.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
- Division of Medical Physics in Radiology, German Cancer Research Centre (DKFZ), Heidelberg, Germany.
- Swiss Center for Musculoskeletal Imaging (SCMI), Zurich, Switzerland.
- Swiss Innovation Hub, Siemens Healthineers International AG, Zurich, Switzerland.
- Siemens Healthineers AG, Erlangen, Germany.
- Department Artificial Intelligence in Biomedical Engineering (AIBE), Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
ORCID-Profile: Matthias Rohe, Katharina Tkotz, Armin M Nagel, Stefan Sommer, Florian Knoll, Michael Uder, Nico Egger, Frederik B Laun, Tobias Wilferth
Förderung
- Interdisciplinary Center for Clinical Research (IZKF)
- German Research Foundation (DFG)
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Rohe, M., Tkotz, K., Nagel, A. M., Sommer, S., Brockmüller, M., Knoll, F., Uder, M., Egger, N., Laun, F. B., & Wilferth, T. (2026). Fat/Water Separation at 7 T Using a 3D Radial Sequence With Quasi‐Continuous Echo Times. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70323
Vancouver
Rohe M, Tkotz K, Nagel AM, Sommer S, Brockmüller M, Knoll F, et al. Fat/Water Separation at 7 T Using a 3D Radial Sequence With Quasi‐Continuous Echo Times. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70323
BibTeX
@article{Rohe2026,
author = {Rohe, Matthias and Tkotz, Katharina and Nagel, Armin M and Sommer, Stefan and Brockmüller, Max and Knoll, Florian and Uder, Michael and Egger, Nico and Laun, Frederik B and Wilferth, Tobias},
title = {Fat/Water Separation at 7 T Using a 3D Radial Sequence With Quasi‐Continuous Echo Times},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70323}
}
EUVIMED