Comparative Performance Analysis of an Open Source Motion‐Compensated Cardiac Diffusion Sequence Across Different MRI Systems and Gradient Strengths—A Pilot Study
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70527 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Eric Arbes und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Medical Physics, Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine - Un…. Hauptthemen laut MEDLINE: Image Enhancement, Heart, Artifacts, Image Interpretation, Computer-Assisted. Die Arbeit nennt 62 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Purpose
To validate the performance of an open‐source cardiac diffusion sequence framework and to investigate hardware‐based performance differences.
Methods
Three second order motion compensated sequence files were compiled using Pulseq and applied across different MRI systems to image five volunteers. DTI and SNR analysis was performed to investigate protocol efficacy.
Results
The sequences performed well at the task of imaging short axis views of the myocardium and producing data suitable for diffusion tensor imaging analysis. Significant differences in sequence performance across different systems and within the same system were revealed. Relative SNR utilizing the strongest available gradient system yielded up to 80% more SNR than the weakest tested system. Relative SNR obtained from the same sequence at different scanners differed by up to 42%, suggesting hardware impacts that warrant further investigation.
Conclusion
The presented sequence framework allows for an easy starting point into cardiac diffusion tensor imaging and opens the door to further analysis of hardware impacts on performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Medical Physics, Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Faculty of Medicine - University of Freiburg, Freiburg, Germany.
- Department of Radiology, Stanford University, Stanford, California, USA.
- Department of Cardiology and Angiology, University Heart Center Freiburg-Bad Krozingen, Faculty of Medicine - University of Freiburg, Freiburg, Germany.
ORCID-Profile: Eric Arbes, Ariel Hannum, Johannes Fischer, Qingping Chen, Sebastian Littin, Alexander Maier, Michael Bock, Simon Reiss
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Arbes, E., Hannum, A., Fischer, J., Chen, Q., Littin, S., Maier, A., Bock, M., & Reiss, S. (2026). Comparative Performance Analysis of an Open Source Motion‐Compensated Cardiac Diffusion Sequence Across Different MRI Systems and Gradient Strengths—A Pilot Study. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70527
Vancouver
Arbes E, Hannum A, Fischer J, Chen Q, Littin S, Maier A, et al. Comparative Performance Analysis of an Open Source Motion‐Compensated Cardiac Diffusion Sequence Across Different MRI Systems and Gradient Strengths—A Pilot Study. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70527
BibTeX
@article{Arbes2026,
author = {Arbes, Eric and Hannum, Ariel and Fischer, Johannes and Chen, Qingping and Littin, Sebastian and Maier, Alexander and Bock, Michael and Reiss, Simon},
title = {Comparative Performance Analysis of an Open Source Motion‐Compensated Cardiac Diffusion Sequence Across Different MRI Systems and Gradient Strengths—A Pilot Study},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70527}
}
EUVIMED