Correcting pulse duration effects in the diffusional kurtosis of the multi‐compartment Kärger model

Jens H Jensen iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30608 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Jens H Jensen. Beteiligte Einrichtung: Center for Biomedical Imaging, Medical University of South Carolina, Charleston, South Carolina, USA. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Algorithms, Image Interpretation, Computer-Assisted, Artifacts. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIA NIH HHS, Litwin Foundation, National Institute on Aging.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To demonstrate a method of reducing pulse duration effects for the diffusional kurtosis of the multi‐compartment Kärger model (KM) as estimated with a Stejskal‐Tanner DWI sequence. Theory and Methods An effective diffusion time is introduced that corrects errors in the apparent diffusional kurtosis arising from a nonzero pulse duration δ for the multi‐compartment KM. The correction is exact to first order in the diffusion time Δ, and numerical calculations are used to assess how well it reduces pulse duration effects. Specifically, for the two‐compartment KM, the deviations of the apparent kurtosis obtained with the Stejskal‐Tanner sequence from the exact kurtosis are calculated for the full range of δ and Δ, and similar calculations are performed for the deviation in the derivative of the kurtosis with respect to Δ. For the general multi‐compartment KM, upper bounds on the maximum magnitude of the deviations are determined. Application of the correction to estimation of intercompartmental exchange rates is illustrated with several examples.

Results

For the two‐compartment KM, the correction reduces the deviation of the apparent kurtosis and its time derivative for most values of δ and Δ. For the general multi‐compartment KM, the maximum deviation magnitude, relative to the initial kurtosis, is 2.26% for the uncorrected kurtosis and 0.57% after correction. The correction reduces the maximum deviation magnitude of the derivative from 46% to less than 1%.

Conclusion

Pulse duration effects for the kurtosis of the multi‐compartment KM can be strongly suppressed by applying the effective diffusion time correction.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diffusion Magnetic Resonance ImagingAlgorithmsImage Interpretation, Computer-AssistedArtifactsImage EnhancementBrainSignal Processing, Computer-AssistedHumansReproducibility of ResultsComputer SimulationSensitivity and Specificity

Stichwörter der Autor:innen

Kärger modelbraindiffusionkurtosispulse durationwater exchange

Autor:innen und Institutionen

  • Center for Biomedical Imaging, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Department of Neuroscience, Medical University of South Carolina, Charleston, South Carolina, USA.
  • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, South Carolina, USA.

ORCID-Profile: Jens H Jensen

Förderung

  • NIA NIH HHS
  • Litwin Foundation
  • National Institute on Aging

Zitieren

APA

Jensen, J. H. (2025). Correcting pulse duration effects in the diffusional kurtosis of the multi‐compartment Kärger model. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30608

Vancouver

Jensen JH. Correcting pulse duration effects in the diffusional kurtosis of the multi‐compartment Kärger model. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30608

BibTeX
@article{Jensen2025,
  author = {Jensen, Jens H},
  title = {Correcting pulse duration effects in the diffusional kurtosis of the multi‐compartment Kärger model},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30608}
}
Kostenlos starten →