Timing Is Everything: How Subtle Timing Changes in MRI Echo Planar Imaging Can Significantly Alter Mechanical Vibrations and Sound Level
Magnetic Resonance in Medicine · 2026-06 · 95 (6) · 3116-3130
DOI: 10.1002/mrm.70271 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 01.02.2026. Verfasst von Amir Seginer und 6 weiteren Autor:innen. Beteiligte Einrichtung: Life Sciences Core Facilities, Weizmann Institute of Science, Rehovot, Israel. Hauptthemen laut MEDLINE: Echo-Planar Imaging, Artifacts, Image Enhancement, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIH HHS, NIH.
Abstract (Originalsprache)
Purpose
Echo‐Planar Imaging (EPI) is central to fMRI, diffusion MRI, and many dynamic clinical applications, yet rapid gradient switching induces strong mechanical vibrations, generates acoustic noise, and contributes to ghosting artifacts—effects that intensify at ultra‐high fields. This study aims to predict how subtle timing changes in multi‐train EPI can modulate acoustic energy and reduce ghosting without altering reconstruction or acquisition design.
Methods
We derived an analytic model describing the interference between short gradient trains and combined it with the system's acoustic transfer function to predict the acoustic energy of a scan. The model predicts cyclic variations in acoustic power resulting from small timing adjustments of slices and echoes. Validation was performed on a clinical 7 T and an investigational 10.5 T MRI systems by measuring acoustic output and ghosting‐artifact levels while systematically sweeping timing conditions in single and multi‐echo EPI protocols.
Results
Sub‐millisecond timing modifications produced large, predictable changes in the acoustic power; changes ranging from 2‐fold to 47‐fold. Intriguingly, under certain conditions, doubling the acquisitions per unit‐time reduced the minimal acoustic energy two‐fold. In addition, navigator time shifting reduced the ghosting‐artifact intensity, up to 5‐fold, exhibiting cyclic dependence corresponding to the dominant acoustic resonance frequency. These trends were consistent across both the 7 T and the 10.5 T systems.
Conclusion
Minimal timing adjustments provide an effective, reconstruction‐agnostic strategy to reduce acoustic noise, to mitigate ghosting artifacts, and to expand feasible EPI parameters space. The proposed model paves the way towards quieter, higher‐quality EPI scans, particularly valuable for ultra‐high‐field and high‐speed imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Life Sciences Core Facilities, Weizmann Institute of Science, Rehovot, Israel.
- The Azrieli National Institute for Human Brain Imaging and Research, Weizmann Institute of Science, Rehovot, Israel.
- Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA.
- Department of Brain Sciences, Weizmann Institute of Science, Rehovot, Israel.
ORCID-Profile: Amir Seginer, Shahar Goren, Noam Harel, Essa Yacoub, Rita Schmidt
Förderung
- NIH HHS
- NIH
Zitieren
APA
Seginer, A., Bratch, A., Goren, S., Furman-Haran, E., Harel, N., Yacoub, E., & Schmidt, R. (2026). Timing Is Everything: How Subtle Timing Changes in MRI Echo Planar Imaging Can Significantly Alter Mechanical Vibrations and Sound Level. Magnetic Resonance in Medicine, 95(6), 3116-3130. https://doi.org/10.1002/mrm.70271
Vancouver
Seginer A, Bratch A, Goren S, Furman-Haran E, Harel N, Yacoub E, et al. Timing Is Everything: How Subtle Timing Changes in MRI Echo Planar Imaging Can Significantly Alter Mechanical Vibrations and Sound Level. Magnetic Resonance in Medicine. 2026;95(6):3116-3130. doi:10.1002/mrm.70271
BibTeX
@article{Seginer2026,
author = {Seginer, Amir and Bratch, Alexander and Goren, Shahar and Furman-Haran, Edna and Harel, Noam and Yacoub, Essa and Schmidt, Rita},
title = {Timing Is Everything: How Subtle Timing Changes in MRI Echo Planar Imaging Can Significantly Alter Mechanical Vibrations and Sound Level},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {95},
number = {6},
pages = {3116--3130},
doi = {10.1002/mrm.70271}
}