Quantifying Skin Sodium Concentration With Novel Very Rapid Biexponential T 2 Relaxometry and Signal Loss Compensation Methodology
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70506 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Gurparkash Singh und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Alberta, Edmonton, Alberta, Canada. Hauptthemen laut MEDLINE: Sodium, Skin, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada, Canada Research Chairs.
Abstract (Originalsprache)
Purpose
To accurately measure skin tissue sodium concentration (TSC) through development of spin‐3/2 relaxometry methodology that also considers blood sodium.
Methods
M xy generation during RF excitation is substantially reduced by very rapid T 2f . Compensation for TSC measurement requires knowledge of both T 2f and B 1 , but the effects of these two parameters are correlated together with a blood sodium contribution that does not exhibit very rapid T 2f . The new spin‐3/2 relaxometry methodology uses 7 scans with 90° RF pulse durations ranging from 70 to 900 μs (constant TE = 2 ms), 2 scans with 60° and 120° flip‐angles, and 2 scans with TE = 6 and 10 ms to disentangle and simultaneously fit skin T 2f , T 2s , relative B 1 and blood sodium fraction. Simulated M xy from the fitted parameters was used for image compensation to attain skin sodium concentration. A gellan gum phantom with known sodium concentration was used for validation. Calf skin of 20 healthy volunteers was scanned with a surface‐coil at 4.7T, measuring T 2f , T 2s , blood sodium fraction and TSC.
Results
The new methodology measured skin T 2f = 0.10 ± 0.03 ms, T 2s = 13.7 ± 1.5 ms, blood sodium fraction = 0.08 ± 0.05, and ultimately TSC = 90 ± 8 mM, which was not correlated with age or sex. However, blood sodium fraction in this model positively correlated with age.
Conclusion
This study yields the first skin TSC measurements consistent with atomic absorption spectroscopy. The potentially large impact of blood sodium in 23 Na MRI of skin, which may also reflect other fluid‐like space, is a particular concern if single images are used for concentration measurement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Alberta, Edmonton, Alberta, Canada.
- Department of Radiology and Diagnostic Imaging, University of Alberta, Edmonton, Alberta, Canada.
ORCID-Profile: Gurparkash Singh, Rob Stobbe
Förderung
- Natural Sciences and Engineering Research Council of Canada
- Canada Research Chairs
Zitieren
APA
Singh, G., Beaulieu, C., & Stobbe, R. (2026). Quantifying Skin Sodium Concentration With Novel Very Rapid Biexponential T 2 Relaxometry and Signal Loss Compensation Methodology. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70506
Vancouver
Singh G, Beaulieu C, Stobbe R. Quantifying Skin Sodium Concentration With Novel Very Rapid Biexponential T 2 Relaxometry and Signal Loss Compensation Methodology. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70506
BibTeX
@article{Singh2026,
author = {Singh, Gurparkash and Beaulieu, Christian and Stobbe, Rob},
title = {Quantifying Skin Sodium Concentration With Novel Very Rapid Biexponential T 2 Relaxometry and Signal Loss Compensation Methodology},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70506}
}
EUVIMED