First In Vivo 23 Na Human Imaging at 10.5 T Using a Combined Sodium‐Proton Transceiver Body Array
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70201 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Simon Schmidt und 2 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Sodium. Die Arbeit nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, National Institute of Biomedical Imaging and Bioengineering.
Abstract (Originalsprache)
Purpose
To demonstrate the first in vivo human 23 Na MRI at 10.5 T using a novel dual‐tuned transceiver body array and to evaluate a self‐gating approach for respiratory motion compensation, focusing on renal imaging.
Methods
A custom‐built eight‐channel 23 Na‐loop 1 H‐dipole transceiver array was designed, constructed, and characterized. Safe operation was ensured through comparison of electromagnetic simulations against B 1 + and SAR phantom measurements. A programmable motion phantom mimicked respiratory motion to assess self‐gating for both the 23 Na and 1 H acquisitions. Finally, in vivo human abdominal 23 Na and 1 H data were acquired in three healthy volunteers under free breathing, applying the self‐gating approach.
Results
The array's electromagnetic model showed good agreement with experimental data, with B 1 + NRMSE values of 0.16 for 23 Na, 0.32 for 1 H, and SAR NRMSE values of 0.35 for 23 Na and 0.22 for 1 H. Total power limits of 125 and 42 W were implemented for 23 Na and 1 H, respectively. The motion phantom study confirmed that both 23 Na and 1 H self‐gating signals accurately tracked the programmed ground truth motion (correlation coefficients: 0.979 for 23 Na, 0.995 for 1 H). Motion binning significantly improved image sharpness as quantitatively evaluated in the phantom. In vivo, self‐gating robustly captured respiratory motion and effectively enabled motion compensation, as shown in binned images with enhanced anatomical detail.
Conclusions
We successfully performed the first in vivo human 23 Na MRI at 10.5 T. The developed dual‐tuned array and validated self‐gating technique enable 10.5 T imaging in humans, paving the way for quantitative studies of sodium homeostasis and advancing diagnostic capabilities.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, Minnesota, USA.
- Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
ORCID-Profile: Simon Schmidt, Gregory J Metzger
Förderung
- NIBIB NIH HHS
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Schmidt, S., Ertürk, A. M., & Metzger, G. J. (2026). First In Vivo 23 Na Human Imaging at 10.5 T Using a Combined Sodium‐Proton Transceiver Body Array. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70201
Vancouver
Schmidt S, Ertürk AM, Metzger GJ. First In Vivo 23 Na Human Imaging at 10.5 T Using a Combined Sodium‐Proton Transceiver Body Array. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70201
BibTeX
@article{Schmidt2026,
author = {Schmidt, Simon and Ertürk, Arcan M and Metzger, Gregory J},
title = {First In Vivo 23 Na Human Imaging at 10.5 T Using a Combined Sodium‐Proton Transceiver Body Array},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70201}
}
EUVIMED