3D ‐printable phantoms for quantitative dynamic contrast‐enhanced MRI
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30595 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von M Sulaiman Sarwar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Clinical Brain Sciences, University of Edinburgh, Edinburgh, United Kingdom. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Printing, Three-Dimensional, Contrast Media. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, Brain Tumour Charity, Engineering and Physical Sciences Research Council (EPSRC).
Abstract (Originalsprache)
Purpose
A novel 3D‐printed phantom design and methodology are proposed, addressing important requirements for technical validation, quality assurance, and multi‐site harmonization of quantitative DCE‐MRI measurements.
Methods
Phantoms were produced by 3D‐printing (3DP) gels incorporating channels and pores as proxies for blood vessels and extravascular extracellular space, respectively. A flow circuit was designed to reproduce clinically relevant arterial input functions. Using nine gels with variable porosity and channel size, we evaluated the effect of 3DP parameters on DCE‐MRI parameters obtained using the extended Tofts model (ET). Physical gel and fitted model parameters were correlated by multiple linear regression.
Results
All phantoms generated realistic arterial input functions, and tissue‐like signal enhancement curves were accurately modeled by the ET model. As hypothesized, blood plasma volume fraction, v p , was positively associated with the channel volume fraction, v chan , ( B = 0.86, p < 0.001) and showed a weaker, negative association with gel porosity, v pore , ( B = −0.18, p = 0.006). Vascular permeability‐surface area product, PS, was positively associated with both v chan ( B = 0.13 min −1 , p < 0.001); and v pore ( B = 0.051 min −1 , p < 0.001). The extravascular extracellular space (EES) volume fraction, v e , was positively associated with v pore ( B = 0.90, p < 0.001) but not v chan . Fitted parameters were reproducible (coefficient of variation 2.1%–3.2%).
Conclusions
Tailorable 3D‐ printed porous gel phantoms generating tissue‐mimicking DCE ‐ MRI signals have the potential to support validation, quality assurance, and multi‐site harmonization of quantitative DCE ‐ MRI .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Centre for Clinical Brain Sciences, University of Edinburgh, Edinburgh, United Kingdom.
- Institute for Bioengineering, University of Edinburgh, Edinburgh, United Kingdom.
- Glasgow Experimental MRI Centre, School of Psychology and Neuroscience, University of Glasgow, Glasgow, United Kingdom.
- Institute for Materials and Processes, University of Edinburgh, Edinburgh, United Kingdom.
- Edinburgh Imaging, Edinburgh, United Kingdom.
ORCID-Profile: Antoine Vallatos, Michael J Thrippleton
Förderung
- Wellcome Trust
- Brain Tumour Charity
- Engineering and Physical Sciences Research Council (EPSRC)
- Clerk Maxwell Cancer Research Fund
Zitieren
APA
Sarwar, M. S., Vallatos, A., Lau, C. H., Waldman, A., Dimartino, S., & Thrippleton, M. J. (2025). 3D ‐printable phantoms for quantitative dynamic contrast‐enhanced MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30595
Vancouver
Sarwar MS, Vallatos A, Lau CH, Waldman A, Dimartino S, Thrippleton MJ. 3D ‐printable phantoms for quantitative dynamic contrast‐enhanced MRI. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30595
BibTeX
@article{Sarwar2025,
author = {Sarwar, M Sulaiman and Vallatos, Antoine and Lau, Cher Hon and Waldman, Adam and Dimartino, Simone and Thrippleton, Michael J},
title = {3D ‐printable phantoms for quantitative dynamic contrast‐enhanced MRI},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30595}
}
EUVIMED