EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access

A concentric-loops coil enhancing RF coil B efficiency and reducing imaging noise compared to traditional single-loop coil

Xin Li, Xiao-Hong Zhu, Wei Chen

Magnetic Resonance Imaging · 2025

DOI: 10.1016/j.mri.2025.110499 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Magnetic Resonance Imaging, 2025. Verfasst von Xin Li und 2 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, MN, USA. Electronic address: li002316@umn.edu. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, NCI NIH HHS, NINDS NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We proposed a simple approach to optimize a concentric-loops RF coil, consisting of an inner primary coil as the driving coil and an outer secondary coil as a passive resonator. By adjusting the tuning capacitor in the secondary loop to raise its self-resonance frequency to be slightly above the operating Larmor frequency, we enabled the preferred operation condition (additive mode), where both secondary and primary loops contributed to the largely enhanced RF coil transmit and receive magnetic fields (B1) within the area enclosed by the primary loop as compared to a single-loop secondary coil. Compared to a single secondary loop coil as control, electromagnetic (EM) simulations and phantom MRS/MRI at 10.5 Tesla (T) 17O (60.6 MHz), 7 T 31P (120.7 MHz), and 1.5 T 1H (63.8 MHz) operating frequencies show that the concentric-loops coil (8 cm diameter for the secondary loop coil) provides 1.2 to 1.4-fold higher B1 and coil detection sensitivity in the near-coil region (0-3 cm). In the far region, it offers slightly enhanced B1 and coil sensitivity, which becomes similar to the control coil at 9 cm away from the coil plane. Additionally, the concentric-loops coil reduces imaging noise by 15-30 %, collectively, resulting in an ∼1.7-fold signal-to-noise ratio (SNR) gain at the near-coil region. The concentric-loops coil with the presented optimization strategy provides an effective and simple coil design with largely improved imaging SNR compared to a conventional single-loop coil for a broad range of MRS/MRI applications across different field strengths.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage EnhancementPhantoms, ImagingEquipment DesignSignal-To-Noise RatioRadio WavesHumansComputer SimulationEquipment Failure AnalysisReproducibility of ResultsSensitivity and Specificity

Stichwörter der Autor:innen

Low fieldNoise reductionRF coilRF transmit and receive magnetic fieldsSNR gainUltrahigh fieldX-nuclei imaging

Autor:innen und Institutionen

  • Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, MN, USA. Electronic address: li002316@umn.edu.
  • Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, MN, USA. Electronic address: zhuxx022@umn.edu.
  • Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, MN, USA. Electronic address: chenx075@umn.edu.

Förderung

  • NIBIB NIH HHS
  • NCI NIH HHS
  • NINDS NIH HHS
  • NIH

Zitieren

APA

Li, X., Zhu, X. H., & Chen, W. (2025). A concentric-loops coil enhancing RF coil B efficiency and reducing imaging noise compared to traditional single-loop coil. Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2025.110499

Vancouver

Li X, Zhu XH, Chen W. A concentric-loops coil enhancing RF coil B efficiency and reducing imaging noise compared to traditional single-loop coil. Magnetic Resonance Imaging. 2025. doi:10.1016/j.mri.2025.110499

BibTeX
@article{Li2025,
  author = {Li, Xin and Zhu, Xiao-Hong and Chen, Wei},
  title = {A concentric-loops coil enhancing RF coil B efficiency and reducing imaging noise compared to traditional single-loop coil},
  journal = {Magnetic Resonance Imaging},
  year = {2025},
  doi = {10.1016/j.mri.2025.110499}
}
Kostenlos starten →