EUVIMEDOpdag EUVIMED
Open access · CC BY-NC 4.0

Active Electromagnetic Interference Suppression for MRI and Proton Resonance Frequency Shift Thermometry During MRI ‐Guided Microwave Ablation

Qing Dai iD, Jason Chiang, Shu-Fu Shih iD, Wenqi Zhou, David S K Lu, Holden H Wu iD

Magnetic Resonance in Medicine · 2026-10 · 96 (4) · 1572-1588

DOI: 10.1002/mrm.70440 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, published online on 2026-06-01. Written by Qing Dai and 5 co-authors. Institution: Department of Radiological Sciences, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California, USA. Major MEDLINE topics: Microwaves, Magnetic Resonance Imaging, Thermometry, Ablation Techniques. 32 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: NIBIB NIH HHS, National Institute of Biomedical Imaging and Bioengineering.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

To develop a software‐based active electromagnetic interference (EMI) suppression (AES) framework on whole‐body multi‐channel MRI systems for reliable MRI and MR thermometry during MRI‐guided microwave ablation (MWA).

Methods

In addition to primary imaging coils, an unloaded standard body array coil was positioned outside the imaging volume to serve as “auxiliary coils” for detecting and characterizing EMI. At each time frame, principal component analysis compressed the auxiliary coil data to extract the dominant EMI pattern. The EMI coupled into the primary coils was then modeled with channel‐specific impulse response functions and subtracted from the primary coil data on a frame‐by‐frame basis. The framework was evaluated through MWA experiments in gel phantoms and in vivo pig models at 3 T with parallel imaging acceleration. Proton resonance frequency shift (PRF) MR thermometry accuracy was assessed using fiber‐optic probes. The PRF‐derived thermal dose prediction was compared with post‐ablation MRI and tissue gross pathology.

Results

AES effectively suppressed EMI‐induced artifacts and substantially improved image quality, yielding signal‐to‐noise ratio enhancement of 40‐fold (gel) and 13‐fold (in vivo) and an EMI suppression rate > 92%. Reliable PRF temperature mapping was enabled, with temporal mean‐absolute‐error of < 1.4°C and < 0.3°C in ablated and non‐heating regions, respectively. Thermal dose estimation matched post‐ablation MRI and gross pathology findings.

Conclusion

The proposed AES framework effectively suppressed ablation‐device‐related EMI and enabled reliable MRI and MR thermometry during MWA. This approach can be integrated into existing reconstruction pipelines to facilitate the broader clinical adoption of MRI‐guided MWA and other interventions affected by EMI.

Subjects

MeSH terms (MEDLINE)

MicrowavesMagnetic Resonance ImagingThermometryAblation TechniquesSurgery, Computer-AssistedAnimalsSwinePhantoms, ImagingProtonsImage Processing, Computer-AssistedArtifactsSoftware

Author keywords

MR thermometryMRI‐guided interventionsactive EMI suppressionmicrowave ablationproton resonance frequency shift

Substances

Protons

Authors and institutions

  • Department of Radiological Sciences, David Geffen School of Medicine, University of California Los Angeles, Los Angeles, California, USA.
  • Department of Bioengineering, University of California Los Angeles, Los Angeles, California, USA.

ORCID profiles: Qing Dai, Shu-Fu Shih, Holden H Wu

Funding

  • NIBIB NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering

Cite

APA

Dai, Q., Chiang, J., Shih, S. F., Zhou, W., Lu, D. S. K., & Wu, H. H. (2026). Active Electromagnetic Interference Suppression for MRI and Proton Resonance Frequency Shift Thermometry During MRI ‐Guided Microwave Ablation. Magnetic Resonance in Medicine, 96(4), 1572-1588. https://doi.org/10.1002/mrm.70440

Vancouver

Dai Q, Chiang J, Shih SF, Zhou W, Lu DSK, Wu HH. Active Electromagnetic Interference Suppression for MRI and Proton Resonance Frequency Shift Thermometry During MRI ‐Guided Microwave Ablation. Magnetic Resonance in Medicine. 2026;96(4):1572-1588. doi:10.1002/mrm.70440

BibTeX
@article{Dai2026,
  author = {Dai, Qing and Chiang, Jason and Shih, Shu-Fu and Zhou, Wenqi and Lu, David S K and Wu, Holden H},
  title = {Active Electromagnetic Interference Suppression for MRI and Proton Resonance Frequency Shift Thermometry During MRI ‐Guided Microwave Ablation},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {4},
  pages = {1572--1588},
  doi = {10.1002/mrm.70440}
}
Start for free →