EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A Subvoxel Correction of Boiling‐Induced Susceptibility Artifacts in Magnetic Resonance Thermometry and Dosimetry for Monitoring Microwave Thermoablation: A Feasibility Study in a Swine Model

Eber Dantas iD, Rebecca Lafont, Pierre Bour, Thibaut Faller iD, Bruno Quesson iD, Helcio R B Orlande iD, Valéry Ozenne iD

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70129 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Eber Dantas und 6 weiteren Autor:innen. Beteiligte Einrichtung: Politécnica/COPPE, Department of Mechanical Engineering, Federal University of Rio de Janeiro-UFRJ, Rio de Janeiro, Brazil. Hauptthemen laut MEDLINE: Microwaves, Magnetic Resonance Imaging, Liver, Thermometry. Die Arbeit nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fundação Carlos Chagas Filho de Amparo à Pesquisa do Estado do Rio de Janeiro, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Agence Nationale de la Recherche: SMART-HEAT(ANR-24-CE92-0073).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Real‐time monitoring of microwave liver ablation (MWA) using MRI thermometry can be hindered by boiling‐induced susceptibility artifacts. These artifacts cause large temperature measurement errors that prevent accurate lesion size prediction. This study proposes a correction methodology based on removing the contribution of this susceptibility artifact using subvoxel sources of susceptibility.

Methods

In vivo microwave ablations ( N = 23) were performed on seven pig livers and the temperature was monitored using the proton resonance frequency shift (PRFS) method. The boiling‐induced artifacts were observed in 11 ablation cases. To validate the proposed methodology, the thermal dose was computed from the corrected temperature maps, and the resulting lesion estimates were compared in three dimensions with ground‐truth lesion segmentations derived from post‐ablation T1w images. A numerical simulation was also conducted to emphasize the need for a finer spatial discretization.

Results

After correction, a better agreement was noticed between thermal dose prediction and lesion size. Over the 11 cases observed with boiling‐induced susceptibility artifacts, the median volumetric Dice, Total Overlap, and False Negative rates changed by 4.8%, 6.2%, and −11.4%.

Conclusion

Comparison with T1w imaging showed improvements in prediction of lesion volume. The proposed methodology takes into account partial volume effects and the contribution of adjacent slices. It was able to simulate a diverse range of deformed dipole‐like artifacts observed in experimental data.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MicrowavesMagnetic Resonance ImagingLiverThermometryImage Processing, Computer-AssistedAblation TechniquesAnimalsArtifactsSwineFeasibility StudiesModels, AnimalAlgorithms

Stichwörter der Autor:innen

MRIboiling artifactmicrowave ablationsusceptibility bubblesthermal dosethermometry

Autor:innen und Institutionen

  • Politécnica/COPPE, Department of Mechanical Engineering, Federal University of Rio de Janeiro-UFRJ, Rio de Janeiro, Brazil.
  • CNRS, CRMSB, UMR 5536, IHU Liryc, Université de Bordeaux, Bordeaux, France.
  • Certis Therapeutics, Pessac, France.

ORCID-Profile: Eber Dantas, Thibaut Faller, Bruno Quesson, Helcio R B Orlande, Valéry Ozenne

Förderung

  • Fundação Carlos Chagas Filho de Amparo à Pesquisa do Estado do Rio de Janeiro
  • Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
  • Agence Nationale de la Recherche: SMART-HEAT(ANR-24-CE92-0073)
  • Université de Bordeaux RRI "IMPACT"
  • Conselho Nacional de Desenvolvimento Científico e Tecnológico
  • Centre National de la Recherche Scientifique (CNRS), MITI interdisciplinary programs
  • CNPQ and FAPERJ
  • Université de Bordeaux

Zitieren

APA

Dantas, E., Lafont, R., Bour, P., Faller, T., Quesson, B., Orlande, H. R. B., & Ozenne, V. (2026). A Subvoxel Correction of Boiling‐Induced Susceptibility Artifacts in Magnetic Resonance Thermometry and Dosimetry for Monitoring Microwave Thermoablation: A Feasibility Study in a Swine Model. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70129

Vancouver

Dantas E, Lafont R, Bour P, Faller T, Quesson B, Orlande HRB, et al. A Subvoxel Correction of Boiling‐Induced Susceptibility Artifacts in Magnetic Resonance Thermometry and Dosimetry for Monitoring Microwave Thermoablation: A Feasibility Study in a Swine Model. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70129

BibTeX
@article{Dantas2026,
  author = {Dantas, Eber and Lafont, Rebecca and Bour, Pierre and Faller, Thibaut and Quesson, Bruno and Orlande, Helcio R B and Ozenne, Valéry},
  title = {A Subvoxel Correction of Boiling‐Induced Susceptibility Artifacts in Magnetic Resonance Thermometry and Dosimetry for Monitoring Microwave Thermoablation: A Feasibility Study in a Swine Model},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70129}
}
Kostenlos starten →