Effect of spatial resolution and heating duration on accuracy of dynamic proton resonance frequency shift temperature measurement with k-space weighted image contrast reconstruction during focused ultrasound

Nicholas Richards, Michael Malmberg, Henrik Odéen, Robb Merrill, Rock Hadley, Dennis L Parker, Allison Payne

Magnetic Resonance Imaging · 2026-09 · 131 · 110679

DOI: 10.1016/j.mri.2026.110679 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2026-09. Verfasst von Nicholas Richards und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Utah, Salt Lake City, UT, United States; Department of Radiology and Imaging Sciences, University…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NCI NIH HHS, NIH HHS, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Single-reference variable flip angle Stack-of-Stars (SR-VFA-SoS) thermometry with KWIC reconstruction provides simultaneous aqueous and fatty tissue temperature monitoring. This work evaluates the effect of k-space weighted image contrast (KWIC) reconstruction of a SR-VFA-SoS sequence on the accuracy of both proton resonance frequency shift (PRFS) and T1-based thermometry with different temporal footprints, spatial resolutions, echo times, and heating durations. METHODS: Magnetic resonance-guided focused ultrasound (MRgFUS) sonications were performed using an asymmetric KWIC-reconstructed single-reference variable flip angle Stack-of-Stars (SR-VFA-SoS) sequence and a validated segmented echo planar imaging (segEPI) thermometry sequence in tissue-mimicking phantoms and an in vivo rabbit model. KWIC windows of projection length 89-377, voxel resolutions of 1.2 mm isotropic and 1.5 × 1.5 × 2.0 mm, three different sets of echo times, and heating durations of 30.2-123.3 s were evaluated. Peak temperature accuracy, calculated cumulative thermal dose (CTD), and temporal lag were assessed. RESULTS: In phantoms and the animal model, significant differences in peak temperatures were observed between the SR-VFA-SoS and segEPI thermometry at 1.5 × 1.5 × 2.0 mm voxel size, but not 1.2-mm isotropic. Temperature discrepancies were independent of echo time. Prolonged heating duration with lower power at the larger voxel size resulted in lower peak temperature error. Asymmetric KWIC-reconstructed SR-VFA-SoS peak temperatures were consistently temporally shifted later than segEPI peak temperatures. CONCLUSION: SR-VFA-SoS thermometry with KWIC reconstruction offers promising temporal resolution for simultaneous aqueous and fatty tissue temperature monitoring during MRgFUS. However, temporal blurring due to KWIC reconstruction results in underestimation and temporal shifting of peak temperatures for large voxel sizes and short heating durations. These observations highlight the need for careful KWIC and MRgFUS parameter optimization.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedAnimalsPhantoms, ImagingRabbitsThermometryProtonsEcho-Planar ImagingTemperatureReproducibility of ResultsAlgorithms

Stichwörter der Autor:innen

Focused ultrasoundInterventionalReconstructionThermometry

Substanzen

Protons

Autor:innen und Institutionen

  • Department of Biomedical Engineering, University of Utah, Salt Lake City, UT, United States; Department of Radiology and Imaging Sciences, University of Utah Health, Salt Lake City, UT, United States. Electronic address…
  • Department of Radiology and Imaging Sciences, University of Utah Health, Salt Lake City, UT, United States.

Förderung

  • NCI NIH HHS
  • NIH HHS
  • NIBIB NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering
  • National Cancer Institute Division of Cancer Epidemiology and Genetics
  • National Institutes of Health Office of Research Infrastructure Programs

Zitieren

APA

Richards, N., Malmberg, M., Odéen, H., Merrill, R., Hadley, R., Parker, D. L., & Payne, A. (2026). Effect of spatial resolution and heating duration on accuracy of dynamic proton resonance frequency shift temperature measurement with k-space weighted image contrast reconstruction during focused ultrasound. Magnetic Resonance Imaging, 131, 110679. https://doi.org/10.1016/j.mri.2026.110679

Vancouver

Richards N, Malmberg M, Odéen H, Merrill R, Hadley R, Parker DL, et al. Effect of spatial resolution and heating duration on accuracy of dynamic proton resonance frequency shift temperature measurement with k-space weighted image contrast reconstruction during focused ultrasound. Magnetic Resonance Imaging. 2026;131:110679. doi:10.1016/j.mri.2026.110679

BibTeX
@article{Richards2026,
  author = {Richards, Nicholas and Malmberg, Michael and Odéen, Henrik and Merrill, Robb and Hadley, Rock and Parker, Dennis L and Payne, Allison},
  title = {Effect of spatial resolution and heating duration on accuracy of dynamic proton resonance frequency shift temperature measurement with k-space weighted image contrast reconstruction during focused ultrasound},
  journal = {Magnetic Resonance Imaging},
  year = {2026},
  volume = {131},
  pages = {110679},
  doi = {10.1016/j.mri.2026.110679}
}
Kostenlos starten →