EPI proton resonant frequency temperature mapping at 0.5T in the brain: Comparison to single‐echo gradient recalled echo
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30373 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Diego F Martinez und 4 weiteren Autor:innen. Beteiligte Einrichtung: The xMR Labs, Department of Physics and Astronomy, Western University, London, Ontario, Canada. Hauptthemen laut MEDLINE: Echo-Planar Imaging, Brain, Thermography, Body Temperature. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Evaluate the use of both single‐echo gradient recalled echo (SE‐GRE) and EPI approaches to creating temperature maps on a mid‐field head‐only scanner, both in vivo and on a tissue mimicking gel.
Methods
Three 2D protocols were investigated (an SE‐GRE, single‐shot EPI, and an averaged single‐shot EPI). The protocols used either a gradient recalled acquisition or an echo planar acquisition, with EPI parameters optimized for the longer at lower field‐strengths. Phantom experiments were conducted to evaluate temperature tracking while cooling, comparing protocol to measurements from an optical fiber thermometer. Studies were performed on a 0.5T head only MR scanner. Temperature stability maps were produced in vivo for the various protocols to evaluate precision.
Results
The use of an EPI protocol for thermometry improved temperature precision in a temperature control phantom and provided an 18% improvement in temperature measurement precision in vivo. Temperature tracking using a fast (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- The xMR Labs, Department of Physics and Astronomy, Western University, London, Ontario, Canada.
- Research and Development, Synaptive Medical, Toronto, Ontario, Canada.
ORCID-Profile: Diego F Martinez
Zitieren
APA
Martinez, D. F., Wiens, C. N., Harris, C. T., Handler, W. B., & Chronik, B. A. (2025). EPI proton resonant frequency temperature mapping at 0.5T in the brain: Comparison to single‐echo gradient recalled echo. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30373
Vancouver
Martinez DF, Wiens CN, Harris CT, Handler WB, Chronik BA. EPI proton resonant frequency temperature mapping at 0.5T in the brain: Comparison to single‐echo gradient recalled echo. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30373
BibTeX
@article{Martinez2025,
author = {Martinez, Diego F and Wiens, Curtis N and Harris, Chad T and Handler, William B and Chronik, Blaine A},
title = {EPI proton resonant frequency temperature mapping at 0.5T in the brain: Comparison to single‐echo gradient recalled echo},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30373}
}