EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

Optimizing Selective RF Pulses for Enhanced Signal Stability in Turbo Spin Echo Using a Differentiable Extended Phase Graph Model

Madison M Augelli iD, Anuj Sharma iD, Mark A Griswold iD, William A Grissom iD

Magnetic Resonance in Medicine · 2026-07 · 96 (1) · 214-226

DOI: 10.1002/mrm.70340 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, published online on 2026-03-12. Written by Madison M Augelli and 3 co-authors. Institution: Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Image Enhancement, Brain, Image Interpretation, Computer-Assisted. 30 references. The full text is freely available (open access, CC BY 4.0). Funded by: National Science Foundation, Siemens Healthineers, NIH HHS.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

To improve slice profile consistency across echo trains in turbo spin echo (TSE) imaging, thereby reducing image blurring and increasing the accuracy of multi echo spin echo mapping.

Methods

Excitation and refocusing RF pulses were optimized for TSE using a differentiable extended phase graph model that incorporates the spinor profiles of the RF pulses to calculate the magnetization slice profile across the echo train. The pulses were optimized using an L‐BFGS algorithm in PyTorch to minimize an error term on the target signal magnitude with singular value regularization to promote similarity. The performance of the optimized pulses was assessed by comparing to time bandwidth‐matched SLR RF pulses. Slice profile consistency was calculated in simulation and in a homogeneous phantom. Images were acquired in vivo to assess blurring artifacts, and measurements were acquired in a NIST phantom and in vivo to assess improvements in accuracy.

Results

The optimized pulses demonstrated superior performance over time bandwidth‐matched SLR pulses, with a 90% reduction in the standard deviation of the normalized integrated signal at each echo. Optimized pulses increased sharpness in vivo at the edges of CSF and veins perpendicular to the phase‐encoded direction, reduced mapping error in the NIST phantom by 91%, and produced more accurate in vivo maps.

Conclusion

The optimization method enables flexible design of RF pulses in echo train pulse sequences with consistent slice profiles, achieving a target signal progression while also maintaining a constant phase and FWHM between echoes.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingImage EnhancementBrainImage Interpretation, Computer-AssistedImage Processing, Computer-AssistedPhantoms, ImagingAlgorithmsHumansRadio WavesReproducibility of ResultsComputer SimulationSignal Processing, Computer-AssistedArtifacts

Author keywords

RF pulse designextended phase graphoptimizationturbo spin echo

Authors and institutions

  • Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.

ORCID profiles: Madison M Augelli, Anuj Sharma, Mark A Griswold, William A Grissom

Funding

  • National Science Foundation
  • Siemens Healthineers
  • NIH HHS
  • National Institutes of Health

Cite

APA

Augelli, M. M., Sharma, A., Griswold, M. A., & Grissom, W. A. (2026). Optimizing Selective RF Pulses for Enhanced Signal Stability in Turbo Spin Echo Using a Differentiable Extended Phase Graph Model. Magnetic Resonance in Medicine, 96(1), 214-226. https://doi.org/10.1002/mrm.70340

Vancouver

Augelli MM, Sharma A, Griswold MA, Grissom WA. Optimizing Selective RF Pulses for Enhanced Signal Stability in Turbo Spin Echo Using a Differentiable Extended Phase Graph Model. Magnetic Resonance in Medicine. 2026;96(1):214-226. doi:10.1002/mrm.70340

BibTeX
@article{Augelli2026,
  author = {Augelli, Madison M and Sharma, Anuj and Griswold, Mark A and Grissom, William A},
  title = {Optimizing Selective RF Pulses for Enhanced Signal Stability in Turbo Spin Echo Using a Differentiable Extended Phase Graph Model},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {1},
  pages = {214--226},
  doi = {10.1002/mrm.70340}
}
Start for free →