Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Jae-Youn Keum und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Intelligent Precision Healthcare Convergence, Sungkyunkwan University, Suwon, Republic of Korea. Hauptthemen laut MEDLINE: Brain, Image Processing, Computer-Assisted, Magnetic Resonance Imaging, Cine, Magnetic Resonance Imaging. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Research Foundation of Korea.
Abstract (Originalsprache)
Purpose
This study aims to address the suppression of oscillation and ghosting appearing in RF‐spoiled gradient‐echo‐based dynamic imaging, which is caused by the pseudo‐steady state (PSS) of magnetization. Theory and Methods We extended the concept of the previous RF phase‐cycle‐adapted averaging method to RF‐spoiled gradient‐echo‐based cine (or 2D line‐scan) imaging by introducing an appropriate intertrial RF phase‐cycle shift while maintaining flexibility in the number of phase‐encoding lines. The proposed method was validated through theoretical analysis based on the PSS magnetization, 2D Bloch simulations, phantom and human brain imaging on a 3 T clinical scanner. Furthermore, its performance was evaluated using spectral analysis and tSNR maps.
Results
In both phantom and human brain images, the proposed method effectively suppressed ghosting artifacts and oscillatory components at specific frequencies. Averaging over multiple trials, whose number is determined by the number of phase increment ( ψ ) rotations required for the PSS magnetization to return to its initial phase, fully eliminated the ghosts and oscillatory components, and even when the number of trials was reduced, it significantly mitigated them. The proposed method provided robust performance in both phantom and human brain images, with improved signal quality as shown in tSNR maps.
Conclusion
In this work, we theoretically investigated the PSS behavior of magnetization in RF‐spoiled gradient‐echo‐based cine imaging and demonstrated a method to effectively suppress oscillation and ghosting appearing in RF‐spoiled gradient‐echo‐based cine images. This study is expected to provide a valuable tool for various applications of ultrahigh temporal resolution MRI, including fMRI, utilizing cine imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Intelligent Precision Healthcare Convergence, Sungkyunkwan University, Suwon, Republic of Korea.
- Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Republic of Korea.
ORCID-Profile: Jae-Youn Keum
Förderung
- National Research Foundation of Korea
Zitieren
APA
Keum, J. Y., & Park, J. Y. (2026). Suppression of Oscillation and Ghosting in RF ‐Spoiled Gradient‐Echo‐Based Dynamic Imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70497
Vancouver
Keum JY, Park JY. Suppression of Oscillation and Ghosting in RF ‐Spoiled Gradient‐Echo‐Based Dynamic Imaging. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70497
BibTeX
@article{Keum2026,
author = {Keum, Jae-Youn and Park, Jang-Yeon},
title = {Suppression of Oscillation and Ghosting in RF ‐Spoiled Gradient‐Echo‐Based Dynamic Imaging},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70497}
}
EUVIMED