Improved Accuracy of Multidimensional Spatially‐Selective Saturation RF Pulses for Short‐Repetition‐Time Steady‐State Imaging
Magnetic Resonance in Medicine · 2027-01 · 97 (1) · 250-258
DOI: 10.1002/mrm.70573 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 26.08.2026. Verfasst von Yongli He und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement, Brain, Image Interpretation, Computer-Assisted. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NINDS NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
To introduce and evaluate a novel RF pulse design algorithm for multi‐dimensional spatially tailored excitation that is optimized for spoiled steady‐state gradient echo MRI sequences.
Methods
Existing multi‐dimensional RF pulse design algorithms optimize RF and gradient waveforms using a loss function based on a single RF pulse application. These designs are poorly suited for steady‐state MRI sequences where RF pulses are spaced only tens of milliseconds apart, causing potentially large differences between the steady‐state magnetization and the target pattern. We propose a multidimensional RF pulse design algorithm with a loss function based on the steady‐state magnetization of a spoiled gradient‐echo sequence. Our algorithm, steady‐state AutoDiffPulses (SS‐ADP), extends the previously introduced ADP framework for joint RF and gradient pulse design. We evaluated design accuracy and RF energy in phantom and in vivo brain imaging in inner‐volume (IV) and outer‐volume (OV) saturation experiments. We further evaluated the potential of SS‐ADP to improve reduced FOV (rFOV) imaging.
Results
SS‐ADP produced significantly more accurate steady‐state magnetization patterns compared to a conventional single‐excitation design (ADP), with significantly reduced RF energy, for . Furthermore, utilizing SS‐ADP for outer‐volume suppression (OVS) substantially reduced reconstruction error in accelerated rFOV imaging compared to conventional ADP‐OVS.
Conclusion
A joint RF and gradient pulse design algorithm accounting for steady‐state magnetization evolution significantly improves excitation accuracy and reduces SAR in short‐TR spoiled imaging with .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA.
- Applied Physics Program, University of Michigan, Ann Arbor, Michigan, USA.
- Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
ORCID-Profile: Yongli He, Douglas C Noll, Jon-Fredrik Nielsen
Förderung
- NINDS NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
He, Y., Noll, D. C., & Nielsen, J. F. (2027). Improved Accuracy of Multidimensional Spatially‐Selective Saturation RF Pulses for Short‐Repetition‐Time Steady‐State Imaging. Magnetic Resonance in Medicine, 97(1), 250-258. https://doi.org/10.1002/mrm.70573
Vancouver
He Y, Noll DC, Nielsen JF. Improved Accuracy of Multidimensional Spatially‐Selective Saturation RF Pulses for Short‐Repetition‐Time Steady‐State Imaging. Magnetic Resonance in Medicine. 2027;97(1):250-258. doi:10.1002/mrm.70573
BibTeX
@article{He2027,
author = {He, Yongli and Noll, Douglas C and Nielsen, Jon-Fredrik},
title = {Improved Accuracy of Multidimensional Spatially‐Selective Saturation RF Pulses for Short‐Repetition‐Time Steady‐State Imaging},
journal = {Magnetic Resonance in Medicine},
year = {2027},
volume = {97},
number = {1},
pages = {250--258},
doi = {10.1002/mrm.70573}
}
EUVIMED