Radial TRASE: 2D RF encoding through mechanical rotation and active digital decoupling
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70104 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Christopher J Sedlock und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology, University of Alberta, Edmonton, Alberta, Canada. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement, Signal Processing, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada, Grants, Natural Sciences and Engineering Research Council of Canada.
Abstract (Originalsprache)
Purpose
Two‐dimensional (2D) transmit array spatial encoding (TRASE) previously required four radiofrequency fields; however, interactions between transmit (Tx) array elements caused significant challenges for 2D imaging. Here, we present a low‐cost, 2D radial encoding scheme (Radial TRASE) using a simplified two‐coil array. Theory and Methods The system consists of two B 1 phase gradient coils capable of encoding any one transverse direction. By incremental mechanical rotation over a 90° range, the encoding axis can be changed, allowing a complete radial k‐space acquisition. As a first demonstration, a wrist‐sized coil pair was experimentally verified on a 2‐MHz Halbach magnet, incorporating a static B 0 slice‐selection gradient. Although a high level of isolation is achievable geometrically, for a more robust implementation, we demonstrate the capability of active digital decoupling in eliminating residual coupling through a parallel‐transmit system.
Results
Radial TRASE–encoded images of water phantoms were acquired, achieving a resolution better than 1.67 mm. Rotation of the Tx array was performed during the recovery period, which caused no imaging delays. All acquired images show minimal distortions, indicating the advantage of the simplified Tx array. The active digital decoupling technique is demonstrated to eliminate residual coupled currents, effectively increasing the isolation of the two‐coil array to −50 dB. Sequential axial slice images were demonstrated using a uniform B 0 coil to shift the slice position.
Conclusion
Two‐coil Radial TRASE can encode a 2D slice without rapidly switched B 0 gradients. Compared with previous three‐coil or four‐coil Cartesian TRASE, the design and isolation of the Tx array are significantly simplified.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oncology, University of Alberta, Edmonton, Alberta, Canada.
- M Tech Research Canada Ltd., Edmonton, Alberta, Canada.
ORCID-Profile: Christopher J Sedlock, Jonathan C Sharp
Förderung
- Natural Sciences and Engineering Research Council of Canada, Grants
- Natural Sciences and Engineering Research Council of Canada
Zitieren
APA
Sedlock, C. J., Purchase, A. R., Tomanek, B., & Sharp, J. C. (2026). Radial TRASE: 2D RF encoding through mechanical rotation and active digital decoupling. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70104
Vancouver
Sedlock CJ, Purchase AR, Tomanek B, Sharp JC. Radial TRASE: 2D RF encoding through mechanical rotation and active digital decoupling. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70104
BibTeX
@article{Sedlock2026,
author = {Sedlock, Christopher J and Purchase, Aaron R and Tomanek, Boguslaw and Sharp, Jonathan C},
title = {Radial TRASE: 2D RF encoding through mechanical rotation and active digital decoupling},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70104}
}