An Open‐Source Software Toolbox for Rapid Radiofrequency Coil Design and Evaluation in MRI
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70269 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Ilias I Giannakopoulos und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, The Bernard and Irene Schwartz Center for Biomedical Imaging, New York University Grossman School of Medicine, New York, New…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Software. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 82 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIH HHS, NIBIB NIH HHS, Cancer Prevention and Research Institute of Texas.
Abstract (Originalsprache)
Purpose
Image quality and resolution in MRI are fundamentally constrained by the performance of radiofrequency (RF) coils used to excite the spins and receive the signal. Electromagnetic (EM) simulations are essential for optimizing coil performance. Existing tools are often slow, memory‐intensive, and require expensive licenses. To address these limitations, we introduce a novel, comprehensive, open‐source EM simulation tool for RF coil design in MRI. Theory and Methods Our toolbox consists of four complementary software components: (1) a full‐wave 3D EM solver based on the wire‐surface‐volume integral equation (WSVIE), where tensor decompositions reduce memory usage and accelerate simulations for fine body model resolutions; (2) a reduced‐order model technique enabling patient‐specific coil simulations in minutes; (3) a fully automatic circuit co‐simulator for coil tuning, matching, decoupling, preamplifier decoupling, and detuning; and (4) a numerical EM basis generator that can be used to compute ultimate performance metrics. We demonstrated the toolbox with a series of simulations for different applications at 7 T MRI.
Results
Our proposed reduced‐order model WSVIE solver was 70 times faster compared to a commercial solver for the simulation of a 31‐channel 7 T head coil. Our co‐simulator tuned, matched, and decoupled the array in 30 min, compared to a few days time of manual interactions needed with the commercial package. The average difference between the signal‐to‐noise ratio maps was less than 15%.
Conclusion
The proposed open‐source simulation framework enables fast, memory‐friendly, accurate, and anatomy‐specific RF coil array design, making it a powerful tool for RF coil engineering.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, The Bernard and Irene Schwartz Center for Biomedical Imaging, New York University Grossman School of Medicine, New York, New York, USA.
- Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, Texas, USA.
- Department of Radiology, The Center for Advanced Imaging Innovation and Research (CAI2R), New York University Grossman School of Medicine, New York, New York, USA.
ORCID-Profile: Ilias I Giannakopoulos, Bei Zhang, José E Cruz Serrallés, Ryan Brown, Riccardo Lattanzi
Förderung
- NIH HHS
- NIBIB NIH HHS
- Cancer Prevention and Research Institute of Texas
- National Institutes of Health
Zitieren
APA
Giannakopoulos, I. I., Zhang, B., Cruz Serrallés, J. E., Brown, R., & Lattanzi, R. (2026). An Open‐Source Software Toolbox for Rapid Radiofrequency Coil Design and Evaluation in MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70269
Vancouver
Giannakopoulos II, Zhang B, Cruz Serrallés JE, Brown R, Lattanzi R. An Open‐Source Software Toolbox for Rapid Radiofrequency Coil Design and Evaluation in MRI. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70269
BibTeX
@article{Giannakopoulos2026,
author = {Giannakopoulos, Ilias I and Zhang, Bei and Cruz Serrallés, José E and Brown, Ryan and Lattanzi, Riccardo},
title = {An Open‐Source Software Toolbox for Rapid Radiofrequency Coil Design and Evaluation in MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70269}
}
EUVIMED