A Numerical Alternative to MR Thermometry for Safety Validation of Multi‐Channel RF Transmit Coils
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70329 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Alireza Sadeghi-Tarakameh und 14 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Thermometry. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 68 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NINDS NIH HHS, NCRR NIH HHS.
Abstract (Originalsprache)
Purpose
This study proposes a numerical technique to estimate peak local specific absorption rate (SAR) uncertainty of multi‐channel RF coils during the process of safety validation as an alternative to experimental temperature and electric‐field measurements, and demonstrates its use to enable human studies at 10.5 T.
Methods
To ensure patient safety, SAR limits established under international guidelines must not be exceeded. Predicting SAR on state‐of‐the‐art parallel transmit systems relies on electromagnetic simulations, which require extensive experimental validation. Despite a well‐established validation workflow, SAR prediction errors are unavoidable and must be quantified as a safety margin. While MRT tests are commonly used for this purpose, their technical challenges necessitate an alternative. The proposed technique propagates the error between experimentally and numerically acquired distributions to the uncertainty in simulated peak local SAR using Monte‐Carlo simulations without the need for MRT. This method was validated using a 16‐channel transceiver 10.5 T torso coil, as well as an 8‐channel 10.5 T head coil.
Results
The proposed numerical technique proved more conservative than existing MRT‐based SAR error quantification methods across all tested scenarios. Its application to validate three state‐of‐the‐art head coils (16Tx/32Rx, 16Tx/80Rx, and 16Tx/128Rx) led to regulatory approval for human head imaging and high‐quality diffusion and functional MRI results at 10.5 T.
Conclusion
The proposed technique requires only the experimental acquisition of maps for comparison with simulations, enabling the estimation of SAR prediction uncertainty. This technique was applied to three 16‐channel transmit arrays, each used in conjunction with high‐channel‐count receive arrays for in vivo imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA.
ORCID-Profile: Alireza Sadeghi-Tarakameh, Simon Schmidt, Matt Waks, Russell L Lagore, Edward Auerbach, Gregor Adriany, Gregory J Metzger
Förderung
- NIBIB NIH HHS
- NINDS NIH HHS
- NCRR NIH HHS
- National Institute of Biomedical Imaging and Bioengineering
- National Institute of Neurological Disorders and Stroke
Zitieren
APA
Sadeghi-Tarakameh, A., Schmidt, S., Waks, M., Lagore, R. L., Grant, A., Auerbach, E., DelaBarre, L., Knudsen, L., Lenglet, C., Vizioli, L., Yacoub, E., Adriany, G., Metzger, G. J., Ugurbil, K., & Eryaman, Y. (2026). A Numerical Alternative to MR Thermometry for Safety Validation of Multi‐Channel RF Transmit Coils. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70329
Vancouver
Sadeghi-Tarakameh A, Schmidt S, Waks M, Lagore RL, Grant A, Auerbach E, et al. A Numerical Alternative to MR Thermometry for Safety Validation of Multi‐Channel RF Transmit Coils. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70329
BibTeX
@article{SadeghiTarakameh2026,
author = {Sadeghi-Tarakameh, Alireza and Schmidt, Simon and Waks, Matt and Lagore, Russell L and Grant, Andrea and Auerbach, Edward and DelaBarre, Lance and Knudsen, Lasse and Lenglet, Christophe and Vizioli, Luca and Yacoub, Essa and Adriany, Gregor and Metzger, Gregory J and Ugurbil, Kamil and Eryaman, Yigitcan},
title = {A Numerical Alternative to MR Thermometry for Safety Validation of Multi‐Channel RF Transmit Coils},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70329}
}
EUVIMED