Robust Fat Suppression for High‐Resolution DWI at 5 T Using Slice‐Selection Gradient Modulation and Chemical Shift Encoding
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70229 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Fan Liu und 12 weiteren Autor:innen. Beteiligte Einrichtung: Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Adipose Tissue, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 76 Literaturangaben. Gefördert von: Beijing Natural Science Foundation - Haidian Original Innovation Joint Fund, National Key Research and Development Program of China, The Peking Union Medical College Hospital Talent Cultivation Program (Category D).
Abstract (Originalsprache)
Purpose
Chemical shift encoding (Dixon) can simultaneously remove multiple fat peaks in multi‐shot EPI diffusion‐weighted imaging (ms‐ EPI DWI ) at 3 T with low sensitivity to B 0 inhomogeneity. However, this method is not applicable at 5 T , since increased fat off‐resonance frequencies cause slice position mismatches between different fat peaks. In this work, we propose a two‐step strategy combining slice‐selection gradient modulation ( SSGM ) and Dixon to enhance fat suppression for ms‐ EPI DWI at 5 T.
Methods
In the first step, SSGM adjusts the amplitudes of excitation and refocusing slice‐selection gradients according to the off‐resonance frequencies of methyl/methylene fat peaks, so that these fat slices are excited but not refocused. In the second step, the olefinic fat peak is chemical shift encoded and separated from the diffusion‐weighted water images through a joint water/fat separation algorithm with structured low‐rank regularization. The two‐step strategy was evaluated in the leg, head‐and‐neck, and prostate.
Results
In vivo experiments demonstrated that the Dixon‐only methods cannot simultaneously suppress all fat peaks at 5 T , while this problem was addressed by combining SSGM and Dixon. SSGM showed superior suppression for methyl/methylene fat compared to SPAIR . The following Dixon further removed olefinic fat untouched by SPAIR . Qualitative analysis showed improved overall image quality for all anatomies. Prostate experiments showed that the proposed method is also applicable in reduced FOV acquisitions, high‐resolution (1.6‐mm isotropic) and high b value imaging (2800 s/mm 2 ).
Conclusion
The proposed two‐step strategy improved fat suppression in ms‐ EPI DWI at 5 T , which can potentially enhance whole‐body disease screening and diagnosis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China.
- Department of Radiology, LUMC, C.J. Gorter MRI Center, Leiden, the Netherlands.
- Shanghai United Imaging Healthcare, Shanghai, China.
- Research Institute of Tsinghua University in Shenzhen, Shenzhen, Guangdong, China.
- Tsinghua Shenzhen International Graduate School, Tsinghua University, Shenzhen, Guangdong, China.
- Radiology Department, Peking Union Medical College Hospital, Beijing, China.
ORCID-Profile: Fan Liu, Yiming Dong, Diwei Shi, Hua Guo
Förderung
- Beijing Natural Science Foundation - Haidian Original Innovation Joint Fund
- National Key Research and Development Program of China
- The Peking Union Medical College Hospital Talent Cultivation Program (Category D)
Zitieren
APA
Liu, F., Dong, Y., Tang, W., Liu, S., Chen, S., Li, G., Shi, D., Shao, X., Jiang, Y., Xue, H., Zhang, G., Sun, H., & Guo, H. (2026). Robust Fat Suppression for High‐Resolution DWI at 5 T Using Slice‐Selection Gradient Modulation and Chemical Shift Encoding. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70229
Vancouver
Liu F, Dong Y, Tang W, Liu S, Chen S, Li G, et al. Robust Fat Suppression for High‐Resolution DWI at 5 T Using Slice‐Selection Gradient Modulation and Chemical Shift Encoding. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70229
BibTeX
@article{Liu2026,
author = {Liu, Fan and Dong, Yiming and Tang, Wending and Liu, Simin and Chen, Shuo and Li, Guangqi and Shi, Diwei and Shao, Xin and Jiang, Yuancheng and Xue, Huadan and Zhang, Gumuyang and Sun, Hao and Guo, Hua},
title = {Robust Fat Suppression for High‐Resolution DWI at 5 T Using Slice‐Selection Gradient Modulation and Chemical Shift Encoding},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70229}
}
EUVIMED