Combined ADC and T 2 Mapping in the Prostate Using a 3D Reduced FOV Sequence
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70510 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Yannik Ott und 8 weiteren Autor:innen. Beteiligte Einrichtung: Research Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Prostate, Imaging, Three-Dimensional, Diffusion Magnetic Resonance Imaging, Image Interpretation, Computer-Assisted. Die Arbeit nennt 93 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To develop a 3D reduced FOV sequence for combined ADC and T 2 mapping in the prostate in a single scan.
Methods
A 3D ADC and T 2 mapping reduced FOV acquisition is enabled using T 2 and diffusion preparation modules with slab‐selective tip‐down pulses and magnitude stabilizer gradients. Imaging shots are followed by 2D phase navigators for phase correction of diffusion‐prepared shots and dummy RF pulses for T 2 ‐prepared shots, as well as a constant delay to maintain the steady state of the longitudinal magnetization. Mono‐exponential fitting is used for ADC and T 2 mapping. T 2 mapping accuracy was assessed in a phantom against a single‐echo spin‐echo reference. In vivo, the sequence was compared in the prostate of 11 healthy volunteers against a multi‐echo spin‐echo acquisition, as well as standard and reduced FOV (rFOV) single‐shot EPI (ssEPI).
Results
In the phantom experiments, T 2 values were not significantly different from the reference single‐echo spin‐echo measurements. ADC values in vivo were not significantly different from those obtained using ssEPI. T 2 values in vivo showed a mean bias of −21.7 ms compared to those obtained using multi‐echo spin‐echo with the first echo discarded in the fitting. The proposed sequence achieved similar overall image quality scores to those of ssEPI and rFOV ssEPI, improved perceived distortion scores, but lower perceived SNR scores.
Conclusion
The proposed approach enables 3D reduced FOV, single‐scan ADC, and T 2 mapping with mono‐exponential fitting in the prostate.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- Research and Clinical Translation, Magnetic Resonance, Siemens Healthineers AG, Erlangen, Germany.
- MR Research Collaborations, Siemens Healthcare Ltd, Camberley, UK.
- Siemens Healthcare SAS, Courbevoie, France.
- Research Department of Cancer Imaging, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- Department of Radiology, Guy's and St Thomas' NHS Foundation Trust, London, UK.
ORCID-Profile: Yannik Ott, Sarah McElroy, Raphael Tomi-Tricot, Carolyn Horst, Patricia A Gutierrez, Omar Darwish, Vicky Goh, Radhouene Neji
Zitieren
APA
Ott, Y., McElroy, S., Tomi-Tricot, R., Horst, C., Gutierrez, P. A., Khamar, R., Darwish, O., Goh, V., & Neji, R. (2026). Combined ADC and T 2 Mapping in the Prostate Using a 3D Reduced FOV Sequence. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70510
Vancouver
Ott Y, McElroy S, Tomi-Tricot R, Horst C, Gutierrez PA, Khamar R, et al. Combined ADC and T 2 Mapping in the Prostate Using a 3D Reduced FOV Sequence. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70510
BibTeX
@article{Ott2026,
author = {Ott, Yannik and McElroy, Sarah and Tomi-Tricot, Raphael and Horst, Carolyn and Gutierrez, Patricia A and Khamar, Rahul and Darwish, Omar and Goh, Vicky and Neji, Radhouene},
title = {Combined ADC and T 2 Mapping in the Prostate Using a 3D Reduced FOV Sequence},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70510}
}
EUVIMED