An essentially radiation‐transparent body coil integrated with a patient rotation system for MR‐guided particle therapy
Medical Physics · 2024
DOI: 10.1002/mp.17065 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Kilian A Dietrich und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: German Federal Ministry of Education and Research.
Abstract (Originalsprache)
Background
The pursuit of adaptive radiotherapy using MR imaging for better precision in patient positioning puts stringent demands on the hardware components of the MR scanner. Particularly in particle therapy, the dose distribution and thus the efficacy of the treatment is susceptible to beam attenuation from interfering materials in the irradiation path. This severely limits the usefulness of conventional imaging coils, which contain highly attenuating parts such as capacitors and preamplifiers in an unknown position, and requires development of a dedicated radiofrequency (RF) coil with close consideration of the materials and components used.
Purpose
In MR‐guided radiation therapy in the human torso, imaging coils with a large FOV and homogeneous B 1 field distribution are required for reliable tissue classification. In this work, an imaging coil for MR‐guided particle therapy was developed with minimal ion attenuation while maintaining flexibility in treatment.
Methods
A birdcage coil consisting of nearly radiation‐transparent materials was designed and constructed for a closed‐bore 1.5 T MR system. Additionally, the coil was mounted on a rotatable patient capsule for flexible positioning of the patient relative to the beam. The ion attenuation of the RF coil was investigated in theory and via measurements of the Bragg peak position. To characterize the imaging quality of the RF coil, transmit and receive field distributions were simulated and measured inside a homogeneous tissue‐simulating phantom for various rotation angles of the patient capsule ranging from 0° to 345° in steps of 15°. Furthermore, simulations with a heterogeneous human voxel model were performed to better estimate the effect of real patient loading, and the RF coil was compared to the internal body coil in terms of SNR for a full rotation of the patient capsule.
Results
The RF coil (total water equivalent thickness (WET) ≈ 420 µm, WET of conductor ≈ 210 µm) can be considered to be radiation‐transparent, and a measured transmit power efficiency (B 1 + /) between 0.17 µT/ and 0.26 µT/ could be achieved in a volume (Δz = 216 mm, complete x and y range) for the 24 investigated rotation angles of the patient capsule. Furthermore, homogeneous transmit and receive field distributions were measured and simulated in the transverse, coronal and sagittal planes in a homogeneous phantom and a human voxel model. In addition, the SNR of the radiation‐transparent RF coil varied between 103 and 150, in the volume (Δz = 216 mm) of a homogeneous phantom and surpasses the SNR of the internal body coil for all rotation angles of the patient capsule.
Conclusions
A radiation‐transparent RF coil was developed and built that enables flexible patient to beam positioning via full rotation capability of the RF coil and patient relative to the beam, with results providing promising potential for adaptive MR‐guided particle therapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
- Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- Faculty of Physics, Heidelberg University, Heidelberg, Germany.
- Clinical Cooperation Unit Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- National Center for Radiation Research in Oncology (NCRO), Heidelberg, Germany.
- Heidelberg Institute of Radiation Oncology (HIRO), Heidelberg, Germany.
Förderung
- German Federal Ministry of Education and Research
Zitieren
APA
Dietrich, K. A., Klüter, S., Dinkel, F., Echner, G., Brons, S., Orzada, S., Debus, J., Ladd, M. E., & Platt, T. (2024). An essentially radiation‐transparent body coil integrated with a patient rotation system for MR‐guided particle therapy. Medical Physics. https://doi.org/10.1002/mp.17065
Vancouver
Dietrich KA, Klüter S, Dinkel F, Echner G, Brons S, Orzada S, et al. An essentially radiation‐transparent body coil integrated with a patient rotation system for MR‐guided particle therapy. Medical Physics. 2024. doi:10.1002/mp.17065
BibTeX
@article{Dietrich2024,
author = {Dietrich, Kilian A and Klüter, Sebastian and Dinkel, Fabian and Echner, Gernot and Brons, Stephan and Orzada, Stephan and Debus, Jürgen and Ladd, Mark E and Platt, Tanja},
title = {An essentially radiation‐transparent body coil integrated with a patient rotation system for MR‐guided particle therapy},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17065}
}
EUVIMED