Angle reduction for fast scans in gradient‐coil‐free portable MRI using rotational spatial encoding magnetic fields via an efficient encoding capability evaluation

Jun-Qi Yang iD, Shao Ying Huang iD, Yi-Feng Jiang iD, Wenwei Yu iD

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70043 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Jun-Qi Yang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Science and Engineering, Chiba University, Chiba, Japan. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement, Image Interpretation, Computer-Assisted. Die Arbeit nennt 37 Literaturangaben. Gefördert von: Japan Science and Technology Agency (JST) SPRING (Support for Pioneering Research Initiated by the Next Generation).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The goal of this study is to reduce rotation angles for faster scans in a gradient‐coil‐free portable MRI system, employing rotational spatial encoding magnetic fields (rSEM) with nonlinear gradients. This is achieved by utilizing a novel evaluation metric for encoding capability, thereby making optimization feasible.

Methods

A local k‐space‐based evaluation metric for the encoding capability of rSEM is proposed. The evaluation metric for each subregion of the image is calculated by summing the area covered by the spokes in the local k‐space when the SEM rotates; each spoke corresponds to one angle. This evaluation metric is expected to correlate closely with the image quality of the reconstructed images and is applied for angle selection to accelerate scans in rSEM systems. The greedy algorithm is used for optimization. Both simulations and experiments were conducted to validate the effectiveness of the evaluation method and the selected angle set.

Results

The results demonstrate that the proposed evaluation metric for the encoding capability of rSEM is highly correlated with image quality metrics of the corresponding reconstructed images (correlation coefficient greater than 0.90). Using the angle set selected by the greedy algorithm with the proposed method, higher image quality with reduced angles is achieved, thereby shortening scan time. This further validates the accuracy of the proposed evaluation metric and underscores the necessity of rapid assessment for feasible optimization.

Conclusion

Significant reduction of rotation angles is achieved by selecting rotation angles using the proposed evaluation method. This facilitates fast scans in gradient‐coil‐free portable MRI systems with rotating SEM while maintaining high image quality. This advancement is crucial for the practical application of this type of portable MRI.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage EnhancementImage Interpretation, Computer-AssistedAlgorithmsReproducibility of ResultsHumansPhantoms, ImagingEquipment DesignRotationMagnetic FieldsSensitivity and Specificity

Stichwörter der Autor:innen

Portable MRISEMgradient‐coil‐free MRIk‐spacelocal k‐spacenonlinear gradientsrSEM

Autor:innen und Institutionen

  • Graduate School of Science and Engineering, Chiba University, Chiba, Japan.
  • Engineering Product Development, Singapore University of Technology and Design, Singapore, Singapore.
  • Department of Surgery, National University of Singapore, Singapore, Singapore.
  • Center for Frontier Medical Engineering, Chiba University, Chiba, Japan.

ORCID-Profile: Jun-Qi Yang, Shao Ying Huang, Yi-Feng Jiang, Wenwei Yu

Förderung

  • Japan Science and Technology Agency (JST) SPRING (Support for Pioneering Research Initiated by the Next Generation)

Zitieren

APA

Yang, J. Q., Huang, S. Y., Jiang, Y. F., & Yu, W. (2026). Angle reduction for fast scans in gradient‐coil‐free portable MRI using rotational spatial encoding magnetic fields via an efficient encoding capability evaluation. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70043

Vancouver

Yang JQ, Huang SY, Jiang YF, Yu W. Angle reduction for fast scans in gradient‐coil‐free portable MRI using rotational spatial encoding magnetic fields via an efficient encoding capability evaluation. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70043

BibTeX
@article{Yang2026,
  author = {Yang, Jun-Qi and Huang, Shao Ying and Jiang, Yi-Feng and Yu, Wenwei},
  title = {Angle reduction for fast scans in gradient‐coil‐free portable MRI using rotational spatial encoding magnetic fields via an efficient encoding capability evaluation},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70043}
}
Kostenlos starten →