Enhancing cardiac MRI reliability at 3 T using motion‐adaptive B 0 shimming

Yuheng Huang iD, Archana Vadiraj Malagi iD, Xinqi Li, Xingmin Guan iD, Chia-Chi Yang, Li-Ting Huang, Ziyang Long, Jeremy Zepeda, Xinheng Zhang, Ghazal Yoosefian, Xioaming Bi, Chang Gao and 7 more

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70026 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2026. Written by Yuheng Huang and 18 co-authors. Institution: Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Heart, Image Processing, Computer-Assisted. Cited 1 times so far (Crossref); 34 references. The full text is freely available in PubMed Central. Funded by: NINDS NIH HHS, NHLBI NIH HHS, National Institute of Health Sciences.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

Magnetic susceptibility differences at the heart–lung interface introduce B 0 ‐field inhomogeneities that challenge cardiac MRI at high field strengths (≥ 3 T). Although hardware‐based shimming has advanced, conventional approaches often neglect dynamic variations in thoracic anatomy caused by cardiac and respiratory motion, leading to residual off‐resonance artifacts. This study aims to characterize motion‐induced B 0 ‐field fluctuations in the heart and evaluate a deep learning–enabled motion‐adaptive B 0 shimming pipeline to mitigate them.

Methods

A motion‐resolved B 0 mapping sequence was implemented at 3 T to quantify cardiac and respiratory‐induced B 0 variations. A motion‐adaptive shimming framework was then developed and validated through numerical simulations and human imaging studies. B 0 ‐field homogeneity and T 2 * mapping accuracy were assessed in multiple breath‐hold positions using standard and motion‐adaptive shimming.

Results

Respiratory motion significantly altered myocardial B 0 fields ( p < 0.01), whereas cardiac motion had minimal impact ( p = 0.49). Compared with conventional scanner shimming, motion‐adaptive B 0 shimming yielded significantly improved field uniformity across both inspiratory (post‐shim SD ratio : 0.68 ± 0.10 vs. 0.89 ± 0.11; p < 0.05) and expiratory (0.65 ± 0.16 vs. 0.84 ± 0.20; p < 0.05) breath‐hold states. Corresponding improvements in myocardial T 2 * map homogeneity were observed, with reduced coefficient of variation (0.44 ± 0.19 vs. 0.39 ± 0.22; 0.59 ± 0.30 vs. 0.46 ± 0.21; both p < 0.01).

Conclusion

The proposed motion‐adaptive B 0 shimming approach effectively compensates for respiration‐induced B 0 fluctuations, enhancing field homogeneity and reducing off‐resonance artifacts. This strategy improves the robustness and reproducibility of T 2 * mapping, enabling more reliable high‐field cardiac MRI.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingHeartImage Processing, Computer-AssistedHumansReproducibility of ResultsArtifactsMotionMaleAdultFemaleRespirationDeep LearningAlgorithmsBreath HoldingLungMovement

Author keywords

B0 field mapB0 shimcardiac MRIdeep learning

Authors and institutions

  • Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.
  • Department of Radiology & Imaging Sciences, Indiana University School of Medicine, Indianapolis, Indiana, USA.
  • Bioengineering, University of California, Los Angeles, Los Angeles, California, USA.
  • MRI Research Institute, Department of Radiology, Weill Cornell Medicine, New York, New York, USA.
  • MR R&D Collaborations, Siemens Medical Solutions Inc., Los Angeles, California, USA.
  • Department of Imaging, Cedars-Sinai Medical Center, Los Angeles, California, USA.

ORCID profiles: Yuheng Huang, Archana Vadiraj Malagi, Xingmin Guan, Hui Han

Funding

  • NINDS NIH HHS
  • NHLBI NIH HHS
  • National Institute of Health Sciences

Cite

APA

Huang, Y., Malagi, A. V., Li, X., Guan, X., Yang, C. C., Huang, L. T., Long, Z., Zepeda, J., Zhang, X., Yoosefian, G., Bi, X., Gao, C., Shang, Y., Binesh, N., Lee, H. L., Li, D., Dharmakumar, R., Han, H., & Yang, H. R. (2026). Enhancing cardiac MRI reliability at 3 T using motion‐adaptive B 0 shimming. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70026

Vancouver

Huang Y, Malagi AV, Li X, Guan X, Yang CC, Huang LT, et al. Enhancing cardiac MRI reliability at 3 T using motion‐adaptive B 0 shimming. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70026

BibTeX
@article{Huang2026,
  author = {Huang, Yuheng and Malagi, Archana Vadiraj and Li, Xinqi and Guan, Xingmin and Yang, Chia-Chi and Huang, Li-Ting and Long, Ziyang and Zepeda, Jeremy and Zhang, Xinheng and Yoosefian, Ghazal and Bi, Xioaming and Gao, Chang and Shang, Yun and Binesh, Nader and Lee, Hsu-Lei and Li, Debiao and Dharmakumar, Rohan and Han, Hui and Yang, Hsin-Jung R},
  title = {Enhancing cardiac MRI reliability at 3 T using motion‐adaptive B 0 shimming},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70026}
}
Start for free →