Open-source, customizable phantom for low-field magnetic resonance imaging
Magnetic Resonance Materials in Physics, Biology and Medicine · 25.06.2025 · 38 (4) · 727-739
DOI: 10.1007/s10334-025-01270-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, online erschienen am 25.06.2025. Verfasst von Kalina V Jordanova und 2 weiteren Autor:innen. Beteiligte Einrichtung: Physical Measurement Laboratory, National Institute of Standards and Technology, 325 Broadway, Boulder, CO, 80305, USA. kalinaj@alumni.stanford.edu. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 65 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Standards and Technology, National Research Council.
Abstract (Originalsprache)
Objective
This study aimed to describe important criteria for phantom design, while designing an open-source phantom that uses accessible materials and fabrication processes, and that can be easily reproduced and modified by others in the MRI research community. Materials and methods We enumerate considerations related to designing a phantom based on literature and previous experience. We design and use an open-source phantom on a low-field MRI system. The phantom was 3D printed and assembled, and the imaged samples were made from commonly available materials. T1-weighted and T2-weighted axial and coronal images were acquired at 64 mT, and signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and geometric distortion along one dimension were assessed for each image.
Results
Two iterations of the phantom design were made to improve the construction materials and overall form factor for imaging. T1-weighted and T2-weighted images showed contrast between samples and background. T2-weighted images had an 8-10× increase in SNR and CNR compared to T1-weighted images. Geometric distortion measurements were within one-pixel spacing for all scans.
Discussion
An open-source phantom was created to assess MRI scans at low-field. Future users may modify the phantom to suit their needs. User-designed inserts can be added, allowing for validation of many MRI-related measurements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Physical Measurement Laboratory, National Institute of Standards and Technology, 325 Broadway, Boulder, CO, 80305, USA. kalinaj@alumni.stanford.edu.
- Physical Measurement Laboratory, National Institute of Standards and Technology, 325 Broadway, Boulder, CO, 80305, USA.
Förderung
- National Institute of Standards and Technology
- National Research Council
Zitieren
APA
Jordanova, K. V., Russek, S. E., & Keenan, K. E. (2025). Open-source, customizable phantom for low-field magnetic resonance imaging. Magnetic Resonance Materials in Physics, Biology and Medicine, 38(4), 727-739. https://doi.org/10.1007/s10334-025-01270-2
Vancouver
Jordanova KV, Russek SE, Keenan KE. Open-source, customizable phantom for low-field magnetic resonance imaging. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025;38(4):727-739. doi:10.1007/s10334-025-01270-2
BibTeX
@article{Jordanova2025,
author = {Jordanova, Kalina V and Russek, Stephen E and Keenan, Kathryn E},
title = {Open-source, customizable phantom for low-field magnetic resonance imaging},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2025},
volume = {38},
number = {4},
pages = {727--739},
doi = {10.1007/s10334-025-01270-2}
}