Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30441 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Matthias Serger und 3 weiteren Autor:innen. Beteiligte Einrichtung: MR Physics, German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany. Hauptthemen laut MEDLINE: Echo-Planar Imaging, Image Processing, Computer-Assisted, Magnetic Resonance Imaging. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Horizon 2020 Framework Programme.
Abstract (Originalsprache)
Purpose
MR‐based FID navigators (FIDnavs) do not require gradient pulses and are attractive for prospective motion correction (PMC) due to short acquisition times and high sampling rates. However, accuracy and precision are limited and depend on a separate calibration measurement. Besides FIDnavs, stationary NMR field probes are also capable of measuring local, motion‐induced field changes. In this work, a linear model is calibrated between field probe data and motion parameters analog to FIDnav calibration and both tracking methods are compared and combined for PMC.
Methods
FIDnavs and field probe navigators were implemented in a fast 3D‐EPI sequence and calibrated by a linear model to realignment motion parameters of the 3D‐EPI time series. A workflow was established to correct head motion prospectively by FIDnavs, field probe navigators or a combination of both. Large motions were instructed to test the accuracy and the impact on image quality in EPI data.
Results
In a group of five subjects, FIDnavs demonstrated approximately doubled accuracy and precision in comparison with field probe navigators for large motions, especially nodding motions were tracked less accurately by field probes. A combination of both methods could not improve the accuracy consistently. Motion artifacts in high‐resolution data were reduced similarly by both PMC methods, although artifacts remained due to susceptibility‐induced B0 changes.
Conclusion
Stationary field probe navigators can be calibrated equivalently as FIDnavs and enable rapid PMC of large and fast motions. Although they reveal decreased accuracy, their contrast‐independence facilitates the potential insertion into many sequences.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- MR Physics, German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany.
- Department of Physics & Astronomy, University of Bonn, Bonn, Germany.
ORCID-Profile: Matthias Serger, Rüdiger Stirnberg, Philipp Ehses, Tony Stöcker
Förderung
- Horizon 2020 Framework Programme
Zitieren
APA
Serger, M., Stirnberg, R., Ehses, P., & Stöcker, T. (2025). Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30441
Vancouver
Serger M, Stirnberg R, Ehses P, Stöcker T. Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30441
BibTeX
@article{Serger2025,
author = {Serger, Matthias and Stirnberg, Rüdiger and Ehses, Philipp and Stöcker, Tony},
title = {Rapid prospective motion correction using free induction decay and stationary field probe navigators at 7T},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30441}
}
EUVIMED