Spectrally selective and interleaved water imaging and fat imaging ( siWIFI )
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30366 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Soo Hyun Shin und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California, San Diego, La Jolla, California, USA. Hauptthemen laut MEDLINE: Adipose Tissue, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Body Water. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAMS NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
To develop a novel imaging sequence that independently acquires water and fat images while being inherently insensitive to motion.
Methods
The new sequence, termed spectrally selective and interleaved water imaging and fat imaging (siWIFI), uses a narrow bandwidth RF pulse for selective excitation of water and fat separately. The interleaved acquisition method ensures that the obtained water and fat images are inherently coregistered. A radial sampling strategy further reduces motion‐induced artifacts. Phantoms with lipid concentrations ranging from 0% to 50% were scanned to measure fat fraction. Moreover, healthy volunteers were scanned to assess the in vivo feasibility of fat fraction measurement at the hip, knee, and liver. In vivo fat fraction measurements were compared with those from vendor‐provided iterative decomposition of water and fat with echo asymmetry and least‐squares estimation (IDEAL) scans. Furthermore, a magnetization transfer (MT) preparation module was incorporated to demonstrate the feasibility of simultaneous measurement of fat fraction and MT ratio utilizing the siWIFI framework.
Results
The phantom fat fractions measured by siWIFI showed excellent correlation with lipid concentrations ( R 2 = 0.9995, p < 0.0001). In vivo studies demonstrated that the fat fractions obtained from siWIFI were comparable to those from IDEAL. Additionally, siWIFI demonstrates reduced motion artifacts from pulsatile flow in knee imaging compared to IDEAL scans and exhibits less sensitivity to respiratory motion in liver imaging compared to IDEAL scans without breath‐hold. The knee imaging study demonstrated that MT‐prepared siWIFI is capable of generating fat fraction and MT ratio maps simultaneously.
Conclusion
The proposed siWIFI sequence allows selective water–fat imaging and quantification with reduced motion artifacts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, University of California, San Diego, La Jolla, California, USA.
- Research Service, VA San Diego Healthcare System, La Jolla, California, USA.
- GE Healthcare, San Diego, California, USA.
- Shu Chein-Gene Lay Department of Bioengineering, University of California, San Diego, La Jolla, California, USA.
ORCID-Profile: Soo Hyun Shin, Jiyo S Athertya
Förderung
- NIAMS NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Shin, S. H., Tang, Q., Carl, M., Athertya, J. S., Suprana, A., & Ma, Y. (2025). Spectrally selective and interleaved water imaging and fat imaging ( siWIFI ). Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30366
Vancouver
Shin SH, Tang Q, Carl M, Athertya JS, Suprana A, Ma Y. Spectrally selective and interleaved water imaging and fat imaging ( siWIFI ). Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30366
BibTeX
@article{Shin2025,
author = {Shin, Soo Hyun and Tang, Qingbo and Carl, Michael and Athertya, Jiyo S and Suprana, Arya and Ma, Yajun},
title = {Spectrally selective and interleaved water imaging and fat imaging ( siWIFI )},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30366}
}
EUVIMED